Instruction Manual mini CORI-FLOW ML120
9.17.097G
9
Advanced instructions
Hook-up diagrams
Document 9.17.095
Instruction manual PROFINET interface
Document 9.16.146
Hook-up diagram ML Series instruments PROFINET
Document 9.16.118
Hook-up diagram LAB MBC3 Custom bus & IO configurations
All documents referred to in this manual can be downloaded from
http://www.bronkhorst.com/en/downloads
1.7
Model key
The model key on the product label contains information about the technical properties of the instrument as ordered. The
specific properties can be retrieved with the diagram below.
1.8
Customized I/O options
mini CORI-FLOW ML120
instruments offer several customized input/output functions through pin 5 of the M12 connector
as an option. The required I/O options are factory installed as specified at ordering time, and cannot be changed manually.
The possible configurations are described in the table below. See the hook-up diagram for custom bus and I/O
configurations (document 9.16.118) for an explanation of the codes.
Code
Description
000
Disabled, pin 5 is pulled down to 0 Vdc (default selection)
A1V
0…10 Vdc sourcing output, controller
Analog signal for pump or external valve steering (control signal only).
When the controller output is used for pump or external valve steering (only applicable to mass flow
meters with the controller function enabled), make sure to set parameter
Valve maximum
to 0.3 [A]. For mass
flow controllers, the controller output signal represents the valve actuator current. This output is limited to
a value below 10 Vdc, due to the maximum valve current restriction.