Bronkhorst®
Instruction Manual mini CORI-FLOW ML120
9.17.097G
28
•
Green
Pattern
Time
Indication
off
Continuous
Power-off or program not running
on
Continuous
Normal Operation Mode
short flash
0.1 sec on,
2 sec off
No bus communication, valves are in safe state (PROFIBUS DP, DeviceNet™, EtherCAT®
and PROFINET only). This LED indication is also active when the instrument is in
‘Initialization Mode' (Init Reset = '73')
normal flash
0.2 sec on,
0.2 sec off
Special function mode; the instrument is busy performing a special function (e.g.
auto-zero or self-test)
long flash
2 sec on,
0.1 sec off
Configuration mode; the baud rate and bus type for the 9-pin D-sub side connector
are set to 38400 and RS232 FLOW-BUS (ProPar)
•
Red
Pattern
Time
Indication
off
Continuous
No error
on
Continuous
Critical error; the instrument needs servicing before it can be used
short flash
0.1 sec on,
2 sec off
Fieldbus specific warning message:
FLOW-BUS
Node occupied: re-install instrument
PROFIBUS DP
No data exchange between master and slave (automatic recovery)
Modbus
Data is being received or transmitted
DeviceNet™
Minor communication error
EtherCAT®
Instrument is not in OP mode
PROFINET
No application relation established
normal flash
0.2 sec on,
0.2 sec off
Incorrect fluid set configuration
(see
Digital parameters - Fluid set
; in this case the
valves are in safe state)
AND/OR:
Fieldbus specific warning message:
FLOW-BUS
Waiting for communication, check communication settings of all
FLOW-BUS devices in the fieldbus setup. Usually the ‘last node
address’ setting of one of the devices is incorrect.
PROFIBUS DP
Not used
Modbus
Not used
DeviceNet™
No bus power
EtherCAT®
Not used
PROFINET
Not used
long flash
2 sec on,
0.1 sec off
Fieldbus specific warning message:
FLOW-BUS
Not used
PROFIBUS DP
Requested parameter not available
Modbus
Not used
DeviceNet™
Serious communication error; manual intervention needed
EtherCAT®
Configuration error
PROFINET
Configuration error (e.g. a requested parameter is not available)
•
Green and
•
red (alternating)
Wink
Time
Indication
slow wink
1 sec on,
1 sec off
Alarm indication; minimum/maximum alarm, power-up alarm, limit reached or batch
reached
normal wink
0.2 sec on,
0.2 sec off
Wink mode; by sending a command to the
Wink
parameter, the instrument flashes its
LEDs to indicate its position in a (large) system.
fast wink
0.1 sec on,
Selected action started (after releasing the multifunctional switch)