Bronkhorst®
Instruction Manual mini CORI-FLOW ML120
9.17.097G
24
3.3.2
Software (DDE applications)
Examples of free Bronkhorst® DDE client applications: FlowPlot and FlowView. Other software programs supporting DDE are
for example MS-Office, LabVIEW, InTouch and Wizcon.
Bronkhorst® software applications 'FlowView' (left) and 'FlowPlot' (right)
FlowDDE and other Bronkhorst® applications are available on the support CD or can be downloaded from
http://www.bronkhorst.com/en/downloads
3.3.3
Baud rate, node address and parity
mini CORI-FLOW ML120
series instruments support the following baud rates for RS232 communication. The factory selected
baud rate is indicated on the instrument label. Refer to
Digital parameters - Network configuration
for changing the baud
rate settings for the instrument. The default baud rate for RS232 communication is 38400 Baud.
Interface/medium:
RS232
Protocol:
FLOW-BUS
Baud rate:
9600
16200
38400
57600
115200
230400
460800
Node address:
3
Parity:
None
Make sure that the instrument’s baud rate corresponds with the baud rate of the application the instrument is
communicating with, otherwise no communication can be established.
For RS232 communication, the maximum cable length is 10 m for baud rates up to 38400 Baud. For higher baud rates,
use cable lengths of maximum 3 m.
Changes made to the network settings will
not
be restored by a factory reset.