General information

Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the
time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temper-
atures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used.
Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the
laser marking becoming unreadable visually or by machine for stainless steel.
Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water
used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in
the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water
and then drying.
Additional drying, if necessary.
Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are
compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used
for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Fail-
ure to do so can result in the following problems:

Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the applica-
tion/process solution only needs to be of pH >8 to cause visible surface changes.

Material damage such as corrosion, cracks, fracturing, premature aging or swelling.

Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause
corrosion.

Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-
k-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.

Disassembling the product before carrying out the reprocessing procedure

Disassemble the product immediately after use, as described in the respective instructions for use.

Open up instruments with hinges.

Preparations at the place of use

If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example.

Remove any visible surgical residues to the extent possible with a damp, lint-free cloth.

Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.

Preparation before cleaning

Dismantle the product prior to cleaning,  see Disassembling.

Cleaning/disinfection

Product-specific safety notes on the reprocessing procedure

Immersion treatment in a 3 % H

2

O

2

 solution for approx. 5 minutes is a particularly effective and gentle method to

dissolve encrustations from HF instruments. Subsequently, the debris can be removed by hand, with a medium-hard
brush and or in an ultrasonic bath. This is followed by the conventional reprocessing steps.

Use suitable cleaning/disinfecting agents if the product is put away in a wet condition. To prevent foam forma-
tion and reduced effectiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the
product thoroughly with running water.

Validated cleaning and disinfection procedure

Manual cleaning/disinfection

Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution.

After manual cleaning/disinfection, check visible surfaces visually for residues.

Repeat the cleaning /disinfection process if necessary.

Manual cleaning with immersion disinfection

D–W:

Drinking water

FD–W:

Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)

RT:

Room temperature

*Recommended: BBraun Stabimed

Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.

Phase I

Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.

Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.

If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.

Mobilize non-rigid components, such as set screws, links, etc. during cleaning.

Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.

Phase II

Rinse/flush the product thoroughly (all accessible surfaces) under running water.

Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.

Drain any remaining water fully.

Phase III

Fully immerse the product in the disinfectant solution.

Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.

Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe.
Ensure that all accessible surfaces are moistened.

Phase IV

Rinse/flush the product thoroughly (all accessible surfaces).

Mobilize non-rigid components, such as set screws, joints, etc. during final rinse.

Rinse lumens with an appropriate disposable syringe at least five times.

Drain any remaining water fully.

Phase V

Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning
and disinfection procedure.

Mechanical cleaning/disinfection with manual pre-cleaning

Note

The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).

Note

The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.

Manual pre-cleaning with a brush

D–W:

Drinking water

RT:

Room temperature

*Recommended: BBraun Stabimed

Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.

Phase I

Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.

Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.

If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.

Mobilize non-rigid components, such as set screws, links, etc. during cleaning.

Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.

Phase II

Rinse/flush the product thoroughly (all accessible surfaces) under running water.

Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.

CAUTION

Damage to the product due to inappropriate cleaning/disinfecting agents and/or 
excessive temperatures!

Use cleaning and disinfecting agents approved for plastics and high-grade 
steel according to manufacturer’s instructions.

Observe specifications regarding concentration, temperature and exposure 
time.

Do not exceed the maximum permitted cleaning temperature of 93 °C.

Validated procedure

Specific requirements

Reference

Manual cleaning with immersion 
disinfection

Cleaning brush, e.g., PM995R 
or GK469R

20 ml disposable syringe

When cleaning instruments 
with movable hinges, ensure 
that these are in an open posi-
tion and, if applicable, move 
the joint while cleaning.

Drying phase: Use a lint-free 
cloth or medical compressed air 

Chapter Manual cleaning/disinfec-
tion and sub-chapter:

Chapter Manual cleaning with 
immersion disinfection

Manual pre-cleaning with brush 
and subsequent mechanical alka-
line cleaning and thermal disinfec-
tion

Cleaning brush, e.g., PM995R 
or GK469R

20 ml disposable syringe

Place the instrument in a tray 
that is suitable for cleaning 
(avoiding rinsing blind spots).

Connect components with 
lumens and channels directly to 
the rinsing port of the injector 
carriage.

Keep working ends open for 
cleaning.

Place instruments in the tray 
with their hinges open.

Chapter Mechanical cleaning/dis-
infection with manual pre-clean-
ing and sub-chapter:

Chapter Manual pre-cleaning 
with a brush

Chapter Mechanical alkaline 
cleaning and thermal disinfect-
ing

Phase

Step

T
[°C/°F]

t
[min]

Conc. 
[%]

Water 
quality

Chemical

I

Disinfecting 
cleaning

RT (cold)

>15

2

D–W

Aldehyde-free, phenol-free, 
and QUAT-free concentrate, 
pH ~ 9*

II

Intermediate 
rinse

RT (cold)

1

-

D–W

-

III

Disinfection

RT (cold)

15

2

D–W

Aldehyde-free, phenol-free, 
and QUAT-free concentrate, 
pH ~ 9*

IV

Final rinse

RT (cold)

1

-

FD-W

-

V

Drying

RT

-

-

-

-

Phase

Step

T
[°C/°F]

t
[min]

Conc. 
[%]

Water 
quality

Chemical

I

Disinfec-
tant

                      

cleaning

RT (cold)

>15

2

D–W

Aldehyde-free, phenol-free, 
and QUAT-free concentrate, 
pH ~ 9*

II

Rinsing

RT (cold)

1

-

D–W

-

Содержание Aesculap 45989

Страница 1: ...ische beschrijving Bipolaire demonteerbare buisschachtinstrumenten Bruksanvisning Teknisk beskrivning Bipol ra demonteringsbara r rskaftinstrument N vod k pou it Technick popis Bipol rn rozlo iteln n...

Страница 2: ...of the cut Note To ensure effective coagulation make certain e g by setting the scissor blades at various angles to the tissue that the tissue cut by the metal ceramic scissor blade is in contact wit...

Страница 3: ...an appropriate cleaning brush for at least 1 min Mobilize non rigid components such as set screws links etc during cleaning Thoroughly rinse through these components with the cleaning disinfectant so...

Страница 4: ...nd has the data on file The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products Other sterilization cycles may also be suita...

Страница 5: ...wird desto gr er wird die Koagulationszone am Schnittrand Hinweis Darauf achten dass das Gewebe z B durch Schr nkung der Scherenbl tter zum Gewebe am Metall Keramikscheren blatt auch den Metallteil b...

Страница 6: ...eeigneten Reinigungsb rste durch b rsten Nicht starre Komponenten wie z B Stellschrauben Gelenke etc bei der Reinigung bewegen Anschlie end diese Stellen mit der reinigungsaktiven Desinfektionsl sung...

Страница 7: ...aktionierten Vakuumverfahren bei 134 C Haltezeit 5 min Bei gleichzeitiger Sterilisation von mehreren Produkten in einem Dampfsterilisator Sicherstellen dass die maxi mal zul ssige Beladung des Dampfst...

Страница 8: ...ra la zone de coagulation au bord de la section Remarque Veiller ce faisant ce que le tissu soit galement au contact de la partie m tallique de la lame de ciseau c ramique m tal p ex en inclinant les...

Страница 9: ...ettoyage appropri e jusqu ce qu aucun r sidu ne soit plus visible sur la surface Le cas ch ant brosser les surfaces non visibles pendant au moins 1 min avec une brosse de nettoyage appro pri e Pendant...

Страница 10: ...5 min En cas de st rilisation simultan e de plusieurs produits dans un st rilisateur vapeur veiller ce que le charge ment maximal autoris du st rilisateur vapeur indiqu par le fabricant ne soit pas d...

Страница 11: ...tamente se realice la secci n mayor ser la zona coagulada en los bordes del corte Nota Asegurarse de que el tejido tambi n toca la parte met lica de la hoja de cer mica y metal por ejemplo triscando l...

Страница 12: ...e al menos 15 min Com probar que todas las superficies accesibles quedan humedecidas Limpiar el producto con un cepillo de limpieza adecuado hasta eliminar todos los restos de la superficie Cuando pro...

Страница 13: ...134 C durante 5 min Si se esterilizan varios productos al mismo tiempo en un esterilizador a vapor Asegurarse de que no se sobrepasa la carga m xima del esterilizador a vapor permitida por el fabrican...

Страница 14: ...elle forbici e poi procedendo al taglio attivando lentamente la corrente HF Quanto pi lentamente si effettua il taglio tanto pi grande diviene la zona di coagulazione sul bordo di taglio Nota Accertar...

Страница 15: ...no per pulizia idoneo nella soluzione finch sulla superficie non pi ricono scibile alcun residuo Se necessario spazzolare le superfici non visibili per almeno 1 min con uno spazzolino per pulizia idon...

Страница 16: ...n Per la sterilizzazione contemporanea di pi prodotti in una sterilizzatrice a vapore accertarsi che non venga superato il carico massimo ammesso per la sterilizzatrice secondo le indicazioni del prod...

Страница 17: ...lentamente o tecido activando simultaneamente a corrente AF Quanto mais lento for o corte maior ser a zona coagulada nos bordos do corte Nota Preste aten o para que no caso de l minas de cer mica e m...

Страница 18: ...acess veis ficam molhadas Limpar o produto com uma escova adequada na solu o at os res duos serem completamente removidos da superf cie Se aplic vel limpar as superf cies n o vis veis durante pelo men...

Страница 19: ...de 5 min No caso de esteriliza o simult nea de v rios produtos num esterilizador a vapor assegurar que a carga m xima admiss vel do esterilizador a vapor definida pelo fabricante n o excedida Armazena...

Страница 20: ...atsen ook het metalen deel van het metaal keramisch schaarblad aanraakt aangezien er anders geen coagulatie plaatsvindt zie Fig 2 Grijpvarianten De bipolaire greep A kan zowel in de traditionele ringg...

Страница 21: ...nminste 1 min met een geschikte reini gingsborstel Beweeg alle bewegende onderdelen zoals stelschroeven scharnieren enz tijdens de reiniging Spoel deze plekken vervolgens minstens 5 maal grondig door...

Страница 22: ...134 C verblijftijd 5 min Wanneer meerdere producten tegelijk worden gesteriliseerd in de stoomsterilisator Let erop dat de maximale belading van de stoomsterilisator die de fabrikant opgeeft niet word...

Страница 23: ...e blir koagulationszonen p snittets kant Tips Se till att v vnaden ocks ber r metalldelen p metall keramiksaxbladet t ex genom skr nkning av saxbladen mot v vnaden eftersom det annars inte sker n gon...

Страница 24: ...erskruvar leder etc vid reng ringen Spola d refter igenom dessa st llen grundligt minst 5 g nger med den aktivt reng rande desinfektionsl sningen och en passande eng ngsspruta Fas II Sk lj av igenom p...

Страница 25: ...aktionerade vakuummetoden vid 134 C i 5 minuter Om flera produkter steriliseras samtidigt i en ngsterilisator Se till att maximal till ten last i ngsterilisatorn enligt tillverkarens anvisningar inte...

Страница 26: ...5 6 7 8 9 10 11 A B Aesculap Aesculap 1 000 Vp IEC DIN EN 60601 2 2 Fig 2 A Fig 7 7 Fig 3 A B 3 A Fig 4 B 1 Fig 4 B 5 6 4 Fig 1 B 6 4 10 Fig 5 5 6 4 Fig 5 A B 3 A Fig 6 1 B 4 A 3 Fig 6 7 11 4 Fig 8 4...

Страница 27: ...www extranet bbraun com Aesculap 6 45 C VAH FDA CE pH 8 www a k i org Rote Brosch re 6 5 H2O2 BBraun Stabimed I 15 1 5 II III 5 IV 5 V 93 C PM995R GK469R 20 PM995R GK469R 20 T C F t I 15 2 pH 9 II 1 I...

Страница 28: ...ulap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de A 5 6 4 Aesculap Aesculap TA Nr 010096 12 12 V6 nd Nr 45989 T C F t...

Страница 29: ...z na na okraji ezu Upozorn n Dbejte na to aby se tk nap k en m elist n ek k tk n na elisti d ek z kovov ho a keramick ho materi lu dot kala tak kovov sti proto e jinak ke koagulaci nedojte viz Fig 2...

Страница 30: ...u i t n pohybujte Z v rem tato m sta d kladn propl chn te pomoc vhodn st ka ky na jedno pou it dezinfek n m roztokem s ist c m inkem minim ln v ak 5 kr t F ze II V robek d kladn opl chn te propl chn t...

Страница 31: ...n m steriliz toru zajist te aby nebylo p ekro eno maxim ln dovolen nalo en parn ho steriliz toru podle daj v robce Skladov n Steriln v robky skladujte v obalech nepropou t j c ch choroboplodn z rodky...

Страница 32: ...ostrzami no yczek a nast pnie powoli przeci ta przy jednoczesnej aktywacji pr du HF Im d u szy jest czas wykonywania ci cia tym wi ksza strefa koagulacji przy brzegu ci cia Notyfikacja Zwr ci uwag na...

Страница 33: ...roztworem Produkt czy ci w roztworze za pomoc odpowiedniej szczotki tak d ugo a na powierzchni nie b d rozpozna walne pozosta o ci Je eli dotyczy niewidzialne powierzchnie nale y co najmniej 1 przeczy...

Страница 34: ...y z DIN EN 285 i walidowany w oparciu o DIN EN ISO 17665 Sterylizacja metod pr ni frakcjonowanej w temp 134 C czas przetrzymania 5 min W przypadku r wnoczesnej sterylizacji wielu produkt w w jednym st...

Страница 35: ...pri kovovo keramickom ostr no n c napr rozvodom ostria ku tka nivu preto e inak k iadnej koagul cii ned jde pozri Fig 2 chopov varianty Bipol rny chop A sa d stl a tak v tradi nom dr iaku s prstencom...

Страница 36: ...pl chnite istiacim dezinfek n m prostriedkom a vhodnou jednor zovou injek nou strieka kou najmenej 5 kr t F za II Cel n stroj v etky pr stupn povrchy opl chnite prepl chnite pod te cou vodou Pohybliv...

Страница 37: ...v jednom parnom steriliz tore zabezpe te aby nebolo prekro en maxim lne pr pustn naplnenie parn ho steriliz tora pod a dajov v robcu Skladovanie Steriln v robky skladujte v obale tesnom proti z rodkom...

Страница 38: ...n meydana gelmeyece inden dokunun rn makas u lar n n dokuya do ru kar l kl e mek suretiyle metal seramik makas ucunda metal k sma da de mesine dikkat edin bkz Fig 2 Tutma e itleri Bipolar kulp A mekan...

Страница 39: ...ar vidalar eklemler vs temizlik s ras nda hareket ettiriniz Daha sonra bu yerleri temizleyici aktif sol syon ile tek kullan ml k bir enjekt r yard m yla 5 kereden az olmamak zere iyice durulay n z Evr...

Страница 40: ...raksiyonlu vakum y ntemiyle 134 C 5 dakika i lem s resiyle sterilizasyon Bir buhar sterilizat r nde ayn anda birden fazla r n sterilize edilecekse Buhar sterilizat r n n retici bilgile rine g re azami...

Страница 41: ...Aesculap RU Monopolar 8 Ring Grip Hold Hand embrace hold 4 1 6 8 7 6 Switch on 2012 02 TA Nr 010096 12 12 V6 nd Nr 45989...

Отзывы: