18
3
1
2
Operation 2016-03-07
Usage; Operation with EndoSync A.I.
Install File Electrode
Hold the push button down and turn the file back and forth
until it is lines up with the notch and goes all the way in.
Release the button to secure it.
• Never put file in or take them out without pressing the button down. This could damage the chuck. Always hold the button down to put a file in or take it out.
• Use only Ni-Ti or properly designed stainless steel files.
• Be careful not cut your finger when putting files in and taking them out.
• Do not let the cutting part of the file touch the electrode; this will wear it out very quickly.
• Some files cannot be used with this electrode.
• Also the files noted below cannot be used. To use these types of files, do not clip on the electrode and use the motor in manual mode.
• Those with a file diameter of more than 1.2 mm.
• Those with chuck shanks that are not perfectly round.
• Gates-Glidden Drills
• Those that have cutting sections with large diameters such as largo burs.
• Do not use files shanks larger than the ISO standard: Diameter 2.334 to 2.350 mm
• After use, do not fail to take the file out.
• Some files cannot use the built-in electrode to make measurement; always check for conductivity before using a file.
• Make sure the file goes all the way in. Give it a light tug to make sure it is held securely.
• Never use stretched, deformed or damaged files.
• Make sure the screw is tight enough. Otherwise, it might come out and be swallowed. Also, measurements might not be accurate.