Operation 2016-03-07
Components Disinfected with Ethanol : Motor Handpiece, Charger, Power Cord, Transmission Cable
Dampen a piece of gauze with ethanol, wring it out and then wipe these components with it.
* The “™” mark indicates that each trade name is a trademark or registered trademark owned by the manufacturer in US or
other territories.
Never wipe components with any solution other than Ethanol for Disinfection (Ethanol 70 to 80 vol%). Other solutions could cause cracking and dis-
coloration.
Never wipe components with a piece of gauze that is excessively wet with Ethanol for Disinfection (Ethanol 70 to 80 vol%). Do not apply or spray
with any fluid. Also, do not immerse in any fluid or wash with water. It could seep inside the instrument and damage it. Be especially careful
around the connection jacks for the transmission cable.
Avoid spilling chemical solutions used for treatment on the motor handpiece, charger, contra angle or any other components. These chemicals could
damage, deform or discolor plastic and metal. Use extra caution to avoid spilling formalin cresol (FC) and sodium hypochlorite as they are quite
strong. Wipe up any chemical spills immediately. (Some chemicals may leave traces even if wiped up immediately.)
Use only Ethanol for Disinfection (Ethanol 70 to 80 vol%) and OPTI-CIDE-3™ Surface Wipes for cleaning. Any other cleaning chemical or products
should not be used including but not limited to the following cleaning products and similar cleaning products listed below because of the potential
damage to the plastic components of the EndoSync.
Take the contra angle out of the paper cup and wipe off any excess oil which may have seeped out.
Dampen a piece of gauze with ethanol, wring it out and then wipe the contra angle with it.
Put the contra angle in a sterilization pouch and autoclave it.
Do not use anything except Ethanol for Disinfection (Ethanol 70 to 80 vol%) for cleaning. Never wipe the contra angle with solutions containing
formalin cresol (FC) or sodium hypochlorite, which damage plastic; wipe them off immediately if they accidentally get on the contra angle.
It is highly recommended that instruments be autoclaved in a sterilization pouch (wrapped) or similar device.
Do not immerse in any fluid or wash with water.
Wipe with Ethanol
Autoclaving the Contra Angle
• CaviWipes™
• CaviCide™
• SANI-CLOTH™