Operation 2016-03-07
Select a memory number (M1 to M6) with the Plus or Minus
switch.
* Before using motor handpiece, use a small hand file, such as
#10 or #15, to penetrate the root canal manually down to the
apical constriction.
* The file electrode must be clipped onto the file for accurate
measurement and instrument control. (In some cases, a root
canal cannot be measured because of an overflow of blood,
saliva or chemicals or because the root canal is blocked.)
Measurement Bar
Meter Numbers
Flash Bar
Current File Tip Position
Memory Bar
(appears only if
set on the Endo-
Sync A.I.)
The meter display appears when the file is inserted in to
the canal. If the Auto Start and Stop is turned ON, the
motor will start running too.
* The numbers 1, 2, and 3 on the meter do not indicate
length in millimeters but are used to estimate how far
the file tip has gone down the canal.
* Press the Select switch to change the display to the
Torque display. Press it again to go back to the canal me-
ter display.
The motor will stop when the file tip reaches the point
specified by the Flash Bar.
A single sustained beep will sound when this happens. If
the unit is set for Apical Reverse, the motor will run back-
wards after it stops.*
If the load on the file exceeds the setting for torque re-
verse, the motor will stop and then reverse its rotation.*
A rapid, repeated three-toned beep will sound when this
happens.
The motor will stop when the file is taken out of the ca-
nal.*
Gradually increase the size of the file until the root canal
preparation is completed.
If necessary, prepare the apical seat.
(*: Depends on setting.)
If the canal is very dry, the Auto Start may not be triggered; in this
case, press the Main switch to start the motor.
• The meter may not appear if the canal is infected or extremely dry. In this case, put a little hydrogen peroxide or saline solution in the canal but do not let it
overflow.
• Accurate measurements cannot be made in some cases because of shape or other conditions. Always check the measurement with an X-ray.
• Do not let the file or metal parts of the contra angle touch the oral mucosa. This could cause the motor to start running and result in injuring the patient.
• An accurate measurement cannot be made if all the connectors are not properly plugged in. If the meter does not move along with the file, stop using the
instrument and check all the connections.