33
Operation 2016-03-07
Problem
Check Points
Response
Meter is not stable dur-
ing use.
• Does the built-in electrode need replace-
ment? Has it been replaced recently?
• Clean and lubricate contra angle.
• Take out the built-in electrode and clean it and the
rotor axle with a brush.
• Replace the built-in electrode.
• Is the screw for the built-in electrode loose?
• Tighten the screw.
Motor handpiece will
not go in reverse rota-
tion.
• Is Torque Reverse setting turned ON?
• Set torque reverse value
• Is Torque Reverse setting too high?
• Reduce torque reverse value
• Is “Apical Action Function” setting turned
OFF?
• Turn “Apical Action Function” setting ON.
• Is the EndoSync A.I. set for Apical Stop?
• Change Apical Stop to Apical Reverse.
Micromotor changes
speed on its own.
• Is Apical Slow Down setting tuned ON?
• When this is turned ON, the motor slows down
as the file approaches the apex. Refer to page
15 for how to turn this setting ON and OFF.
• Is Torque Slow Down setting tuned ON?
• When this is turned ON, the motor slows down
as the torque increases. Refer to page 15 for
how to turn this setting ON and OFF.
Unit turns OFF/ON its
own.
• Was the unit no used for a long time?
• Auto power OFF was probably activated. Press
the Main switch to turn the unit back ON.
• Does “Please Charge” appear in the dis-
play?
• Battery must be charged right away.
• This can happen if the battery is very low
and a large load is applied to the file.
• Battery must be charged right away.
Error 01
• Transmission cable is probably not properly
connected.
• Check cable and connect is properly.
Error 04
• Does this happen repeatedly?
• Clear the memory and restore the initial settings.
Error 06
• Does this happen repeatedly?
• The motor circuits may be malfunctioning. Have
the instrument repaired.