39
Operation 2016-03-07
Essential Performance:
Noise does not substantially change measurement.
Noise will not change operation mode.
Noise will not permanently alter display values.
Transmission Cable:
Length: 1.5 meters
• Use of the parts other than those accompanied or specified by Brasseler USA may result in increased EMC emissions or decreased EMC immunity of the TR-
CM Type BSL.
Recommended separation distances between portable and mobile RF communications equipment and the TR-CM.
The TR-CM is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The cus-
tomer or the user of the TR-CM can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the TR-CM as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz d =1.11
80 MHz to 800 MHz d =
800 MHz to 2.5 GHz d = 2
0.01
0.11
0.1
0.2
0.1
0.35
0.32
0.63
1
1.11
1
2
10
3.51
3.16
6.32
100
11.11
10
20
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec-
tion from structures, objects and people.