15
Operation 2016-03-07
Memory Settings: Additional Operation Settings (Normal mode)
Additional Functions:
Torque Slow Down (TORQ.SL.D.), Linked to canal measurement (APICAL ACT.**),
Apical Reverse or Stop (APICAL ACT.**), Auto Start and Stop (AUTO STT/STP**),
Apical Slow Down (APICAL SL.D.**), Apical Torque Reduction (APICAL TRQ.D.**)
1. Select a memory number for the standby display; press the Plus or Minus switch.
2. Hold down the Select switch for at least 1 second to show the displays for additional opera-
tion settings.
3. Press the Select switch to go from one display to the next.
4. Change the setting; press the Plus or Minus switch.
* The display will go back to the standby display if 5 seconds (initial setting) elapses without
a switch being pressed.
Torque Slow Down:
When this is turned ON, the motor will slow down as the torque load increases.
Standby Display
Apical Torque Reduction
Apical Action Function
**
:
When this is turned ON, Apical Reverse or Stop function be activated.
Apical Reverse or Stop
**
:
The file will reverse or stop when the file tip reaches the Flash Bar.
When Apical Action Function is turned OFF, this display will be skipped.
Auto Start & Stop
**
:
When this is turned ON, the file starts rotating when it is inserted and stops when it is taken
out of the canal.
Apical Slow Down
**
:
When this is turned ON, the file slows down as it approaches the apex.
Apical Torque Reduction
**
:
When this is turned ON, the torque setting that triggers reverse rotation is reduced as the file
tip approaches the apex.
**
These functions are available only when connected to the EndoSync A.I.
Apical Slow Down
Auto Start & Stop
Apical Reverse or Stop
Torque Slow Down
Apical Action Function
Cannot be used along with Apical Torque Reduction function.
If the Torque Reverse Less (TRL) is turned ON, the Torque Slow Down function is disabled.
If the Rotation Mode is set for “Rev”, the Apical Reverse function is disabled.
Cannot be used along with Apical Torque Reduction function.
Cannot be used along with Apical Slow Down or Torque Slow Down functions.
If the Torque Reverse Less (TRL) is turned ON, the Apical Torque Reduction function is disabled.