65
Table 5.
Guidance and Manufacturer’s Declaration—Electromagnetic
Compatibility
(continued)
The LATITUDE Programming System, Model 3300, is suitable for use in
professional health care facility environments. The customer or the user of this
system should assure that it is used in such an environment.
Test
Compliance
Electromagnetic
environment—guidance
1700 - 1900 MHz: 28 V/m
2400 - 2570 MHz: 28 V/m
5100 - 5800 MHz: 9 V/m
Rated power frequency
magnetic field
30 A/m
Electrical fast transients/
bursts
± 2 kV input AC power
± 1 kV SIP/SOP
Surges line-to-line
± 0.5 kV, ± 1 kV input AC
power
Surges line-to-ground
± 0.5 kV, ± 1 kV, ± 2 kV
input AC power
Conducted disturbances
induced by RF fields
3 V/m from 0.15 MHz to 80
MHz
6 V/m in ISM bands from
0.15 MHz to 80 MHz
The ISM bands between
0.15 MHz and 80 MHz are
6.765 MHz to 6.795 MHz
13.553 MHz to 13.567
MHz
26.957 MHz to 27.283
MHz
40.66 MHz to 40.70 MHz.
The amateur radio bands
between 0.15 MHz and 80
MHz are
1.8 MHz to 2.0 MHz
3.5 MHz to 4.0 MHz
5.3 MHz to 5.4 MHz
7.0 MHz to 7.3 MHz
10.1 MHz to 10.15 MHz
14.0 MHz to 14.2 MHz
18.07 MHz to 18.17 MHz
21.0 MHz to 21.4 MHz
24.89 MHz to 24.99 MHz
28.0 MHz to 29.7 MHz
50.0 MHz to 54.0 MHz.