12
The lowest version of USB determines the speed. For example, a USB 3.0
device plugged into a USB 2.0 port runs at 2.0 speed, and a USB 2.0
device plugged into a USB 3.0 port runs at 2.0 speed.
Interrogation and Programming
•
Interrogates and programs the implantable PG
•
Displays records, stores patient data, and allows the physician to evaluate
alternative prescription modes, generate reports, and record episodes
•
Performs tests in an electrophysiology laboratory, in an operating room, in
an emergency room, in clinical environments, or at a patient’s bedside
•
May be used to support diagnostic activities
3
pertaining to implanting,
programming, and monitoring Boston Scientific implantable PGs
•
Provides a Pacing System Analyzer (PSA) application
4
to assess electrical
performance and placement of cardiac lead systems during implant of
cardiac rhythm management devices
•
Provides real-time electronic capture of various events from the PG and
the PSA application
•
Outputs surface ECG and telemetered signals (intracardiac electrograms
and event markers) in PDF format
•
Provides emergency access to STAT SHOCK, PG STAT PACE, PSA STAT
PACE, and DIVERT THERAPY functionality applicable to the PG and PSA
application
•
Provides ZIP telemetry, a cordless, hands-free RF communication option
that allows the Programmer to communicate with the PG
Patient Data Management
The LATITUDE Programming System provides the ability to print, save, or
transfer related data (via Bluetooth
®
or USB pen drive), during or after an
implant/follow-up session, to a clinic computer for processing/transferring data
to external systems (e.g., EMR systems).
Refer to the
Patient Data Management Operator’s Manual (Model 3931)
for
additional information.
Networking
The LATITUDE Programming System provides Ethernet and wireless (Wi-Fi)
connectivity for data transmission. Bluetooth
®
connectivity is available for data
transfer (e.g., to a laptop) and printing.
Refer to the
Network and Connectivity Operator’s Manual (Model 3924)
for
additional networking and connectivity setup and use information.
3.
The LATITUDE Programming System is not intended for use as an ECG monitor or general
diagnostic device.
4.
Refer to the
Pacing System Analyzer (PSA) Operator’s Manual (Model 3222)
for PSA setup and
use information.