7
Before cleaning and disinfecting the Programmer surfaces, power down
the device and disconnect the external power supply. Before operating the
LATITUDE Programming System, let cleaning and disinfection agents
used on the Programmer evaporate.
•
Emissions and interference.
The emissions characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally required),
this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation
measures, such as relocating or reorienting the equipment. Other
equipment may interfere with the LATITUDE Programming System, even
if that equipment complies with the CISPR emission requirements.
•
Lithium-ion battery.
The Model 6753 Battery is a Lithium-ion battery and, as such, is deemed
a Dangerous Good in regards to shipping. Do not return the Model 6753
battery to Boston Scientific. Dispose of the battery as required by local
regulations. If the battery must be shipped, contact your local shipper for
instructions and shipping requirements.
PRECAUTIONS
General
•
Functional impairment due to external damage.
Mechanical impact, for
example dropping the Programmer unpackaged, can permanently impair
the function of the system. Do not use the Programmer if there is apparent
damage. If damage has occurred, contact Boston Scientific to return the
Programmer using the information on the back cover of this manual.
•
Programming System.
Use only the appropriate LATITUDE
Programming System equipped with the appropriate software to program
specific Boston Scientific PGs.
•
Wand use.
For transvenous PG telemetry, use only the Model 6395
Telemetry Wand with the LATITUDE Programming System.
•
Stylus use.
If you want to use a stylus, ensure that it is a projected
capacitance stylus. The use of any other object could damage the
touchscreen.
•
Electrocautery cables.
Keep all electrocautery cables at least 30 cm (12
in) away from the LATITUDE Programming System to avoid false signals
due to electrocautery energy.
•
Leakage current.
Although optional external equipment connected to the
Model 3300 Programmer must meet the applicable leakage-current
requirements for commercial products, it may not meet the more stringent