TROUBLESHOOTING
If the LATITUDE Programming System does not operate properly, check that
electrical cords and cables are securely connected and that cords and cables
are in good working order (i.e., free of visible defects). Possible causes and
corrective actions for problems listed below.
Table 3.
Possible Causes and Corrective Actions for LATITUDE Programming
System Problems
Symptom
Possible Cause
Corrective Action
Telemetry: poor,
intermittent, or no
communication
Incorrect application
software or incorrect
LATITUDE Programming
System for PG
Install proper application
software for PG in use.
Use correct LATITUDE
Programming System for
the PG being interrogated.
Contact Boston Scientific
using the information on
the back cover of this
manual to confirm PG and
Model 3300 Programmer
compatibility.
Incorrect telemetry wand
Use only the Model 6395
Telemetry Wand for
transvenous PGs.
Poor connection between
a telemetry wand and the
Programmer
Disconnect and reconnect
the Model 6395 Telemetry
Wand to the Programmer.
Programmer running on
battery power only
Connect the Programmer
to AC power to improve
telemetry performance.
Excessive radio emissions
from equipment
Reposition the LATITUDE
Programming System.
Also see Noise problems:
ECG.
Incomplete telemetry
communication with Model
6395 Telemetry Wand
Reposition the Model
6395 Telemetry Wand
over the PG; repeat
interrogation.
Flip the wand over.
Disconnect and reconnect
the wand. Turn off the
Programmer, and then
turn it On. Repeat
interrogation.