14
The use of any cables or accessories with the LATITUDE Programming
System other than those provided by or specified by Boston Scientific could
result in increased electromagnetic emissions, decreased electromagnetic
immunity, or electrical shock of the LATITUDE Programming System. Anyone
connecting such cables or accessories to the LATITUDE Programming
System, including the use of MSOs (Multiple Socket Outlets), may be
configuring a medical system and is responsible to ensure that the system
complies with the requirements of IEC/EN 60601-1, Clause 16 for medical
electrical systems.
Optional External Equipment
Optional external equipment can be used with the LATITUDE Programming
System. Contact your Boston Scientific sales representative to determine what
external equipment can be used.
NOTE:
If adding external equipment, you are configuring a medical system
and are responsible to ensure that the system complies with the requirements
of IEC/EN 60601-1, Clause 16 for medical electrical systems.
WARNING:
Do not simultaneously touch the patient and any accessible LATITUDE
Programming System connector or exposed conductor.
CAUTION:
Although optional external equipment connected to the Model
3300 Programmer must meet the applicable leakage-current requirements for
commercial products, it may not meet the more stringent leakage requirements
for medical products. Consequently, all external equipment must be kept
outside the patient environment.
•
Never touch the electrical contacts on the side panels of the Model 3300
Programmer and the patient, a telemetry wand, or any cable at the same
time.
Stand
A stand (Model 6755) is available for the LATITUDE Programming System. It
easily attaches to the bottom of the Programmer with a clip. It provides two
convenient viewing angles and has a storage pocket in the back for storing
cables and wands.
When the stand is used in the flat position, do not use downward force on the
handle as the unit may tip.
To attach the stand, slip the stand under the Programmer and tilt the stand up
to engage the clip as illustrated in Figure 1 Optional Stand for the LATITUDE
Programming System on page 15.