![Boston Scientific Latitude 3300 Скачать руководство пользователя страница 47](http://html.mh-extra.com/html/boston-scientific/latitude-3300/latitude-3300_operators-manual_2796085047.webp)
41
pop-up displays each time you select another part of the Real-time Log Event
display, so that you can use multiple tools anywhere on the display.
The five tools are:
•
Circle tool
–places a circle on the display at the target point.
•
Line tool
–places a dashed vertical line on the display at the target
point.
•
Left scissor tool
–removes the left-hand portion of the display from the
target point.
•
Right scissor tool
–removes the right-hand portion of the display from
the target point.
NOTE:
When using the scissors tools, the original trace is still available
in the Real-time Log along with the actual scissored portion.
•
Annotation tool
- displays a keyboard to type in any notes, which then
appear on the trace.
Electronic Calipers
Use the Electronic Calipers (slide bar) to adjust the time span of the Snapshot
trace. The time interval between the calipers is measured in seconds. A caliper
can be repositioned by selecting it and then dragging it to expand or collapse
the time frame. For detailed instructions on using the Electronic Calipers, refer
to the associated product literature for the PG being interrogated.
Real-time Log Events
PG events that qualify for automatic real-time recording are listed in Table 1 PG
Events on page 41.The device action that initiates storage is recorded in the
Real-time Log.
Table 1.
PG Events
Event Type
Trigger Event
Duration of Recording
(seconds)
Presenting
Initial Interrogation
Completed
12
Electrocautery Mode
Electrocautery Mode
Entered
12
PG STAT PACE
PG STAT PACE
Commanded
12
DIVERT THERAPY
DIVERT THERAPY
Commanded
12
PACE THRESHOLD
TEST (AUTO, A, V, RV,
LV, Ampl, and PW)
Threshold Test Ended
12