![Boston Scientific EKOS Control System 4.0 Скачать руководство пользователя страница 54](http://html.mh-extra.com/html/boston-scientific/ekos-control-system-4-0/ekos-control-system-4-0_user-manual_2796075054.webp)
54
Bost
on Scientific (Mast
er Br
and,
Spir
al Bound Manual T
empla
te
, 8
.5in x 11in Global,
923100
58G),
IFU
, MB,
CU4,
en,
5
1011236
-001A
Black (K) ∆E ≤5.0
Black (K) ∆E ≤5.0
SPECIFICATIONS
Table 23.
Type
Parameter
Value
System
Specifications
Device Frequency
2�05 MHz – 2�35 MHz
Ultrasound Transducer Composition
PZT (Lead Zirconate Titanate)
System Output Power
Automatically controlled, 100 W pulse power maximum 70 W
average device power maximum for both channels (35 W average
device power maximum per channel�)
Maximum Device Temperature
43 °C Temperature Limit
During normal use, ultrasound therapy may cause a temperature
rise in the treatment zone� There are temperature sensors in the
EKOS Device� Control Unit 4�0 monitors the device temperature
and controls power output to limit the device temperature to a
maximum of 43 °C� Brief excursions near or above 43 °C will result
in temporary suspension of ultrasound� The Control Unit will
suspend therapy and alert the operator if the safety limits of 43 °C for
25 seconds or 52 °C for 1 second are exceeded
Power Requirements
100 V – 240 V, 50 Hz/60 Hz, 1�6 A maximum
Physical Dimensions
4�0 in x 10�5 in x 8�3 in (10�2 cm x 26�7 cm x 21�1 cm) (H x W x D)
Environmental
Conditions
Operating Humidity
30 % to 75 %, non-condensing
Operating Temperature
+15 °C to +40 °C
Operating Atmospheric Pressure
73 kPa to 111 kPa
External Storage / Transport Humidity
10 % to 90 %, non-condensing
External Storage / Transport Temperature
-20 °C to +60 °C
External Storage / Transport Atmospheric Pressure
73 kPa to 111 kPa
Table 24. Ordering Information
Product
Catalog Number
EKOS Control Unit model 4�0
H7493959340010
Connector Interface Cable
H7493959340060
Power Cord
Power cord part numbers vary by geography� For the part number for power cord for your geography
please contact your local Boston Scientific representative�
WARRANTY
For device warranty information, visit (www�bostonscientific�com/warranty)�
EKOS and EkoSonic are trademarks of Boston Scientific Corporation or its affiliates�
Sani-Cloth and Super Sani-Cloth are trademarks of Professional Disposables International, Inc�
All other trademarks are the property of their respective owners�
SYMBOL DEFINITIONS
Commonly used medical device symbols that appear on the labeling are defined at www�bostonscientific�com/SymbolsGlossary�
Additional symbols are defined at the end of this document�