Figure 27. Two Channels with Ultrasound Running Screen
2� Monitor the runtime clock to determine how long the channel has delivered ultrasound therapy�
The runtime is shown as Hours:Minutes:Seconds� Time segments not yet reached are dimmed� In the example below, ultrasound therapy has
been running for 23 minutes and 47 seconds (see Figure 28)�
Hours
Minutes
Seconds
00:
23:47
Figure 28. Ultrasound Paused/Ready Screen
Note:
The runtime information is saved for the channel, not for the device connected� If the ultrasound therapy is stopped and restarted, the
timer will resume when the ultrasound therapy is started again� If the device cables are disconnected (e�g�, patient transport) and re-connected
the channel timer will resume when ultrasound therapy is started again� It is strongly recommended that devices be reconnected to the same
channel they started on to ensure the total runtime displayed for the channel reflects the total runtime for the device�
Note:
During normal use, ultrasound therapy may cause a temperature rise in the treatment zone� There are temperature sensors in the
EKOS Device� The Control Unit monitors the device temperature and optimizes the output power to the treatment area�
3� To stop or pause the ultrasound, press the
Stop
button on the Control Unit for the appropriate channel�
When the
Stop
button is pressed, the ready to start instruction is displayed indicating the total elapsed runtime for the ultrasound therapy
(see Figure 29)�