13. Check Mode
It is a mode to check load cell output (mV), check DAC(DAI) output, select option setting, and initialize setting value.
13-1 Operation for each check mode
13-1-1 How to enter check mode
1.
It turns into the setting available status when key is pushed while pushing key is
pushed at measurement status.
2.
Push key in order to move onto next stage at mode selection status.
3.
Mode change and setting value change is available when key or key is pushed at
mode selection status.
4.
Push key to cancel the setting and return to measurement mode for mode cancellation.
At the mode cancelation, the setting until previous stage of cancellation is saved.
It is entered into mode when
turning the power on while
pushing No,2 and No.3 key
under power off status.
Current load cell output voltage is displayed.
000.000mV is displayed under no load and
mV is differently displayed based on dead load
Load cell shall be examined with the display of higher than 015.000mV
or display of ovEr(OVER).
13-1-2 Load cell output voltage check
It is mode to check load cell output voltage. Load cell output voltage is displayed on meter display unit.
Load cell mV voltage can be checked without digital multi-meter. (ex: 003.785mV)
There is an increase in voltage when load is exerted to load cell. Measurement of up to 035.000mV is
available and Over is displayed for value exceeding it. Examine the load cell when displaying load cell
output voltage of 015.000mV or higher. .
Abnormality can be checked by
examining load cell output
voltage.
It moves onto next stage when
pushing No.4 key for 2 sec and
push No.1 key for cancellation.