40

 

Display (small)

RY1 setting value

RY3 setting value

Display (large)

ON

OFF

ON

OFF

ON

OFF

RY2 setting value

L1 (Lo) output

L3 (Hi) output

L2 (OK) output

Display (small)

Display (large)

ON

OFF

L1 (Lo) output

L3 (Hi) output

L2 (OK) output

ON

OFF

ON

OFF

RY1 setting value

RY3 setting value

RY2 setting value

 

    Low limit mode 

Relay output terminal

 

Output condition

 

L1  (LO)  : 

Measurement value   

   RY1 setting value 

L2  (OK)  : 

Measurement value   

    RY2 setting value 

L3  (HI)   : 

Measurement value   

    RY3 setting value 

 

 
 

 
 
 

 
 
 

 
 
 

 
 
 

 
 
 

 

 

 

 

 

 

 
 

 

 

    Low Low High limit mode 

Relay output terminal

 

Output condition

 

L1  (LO)  : 

Measurement value   

   RY1 setting value 

L2  (OK)  : 

Measurement value   

    RY2 setting value 

L3  (HI)   : 

Measurement value   

    RY3 setting value 

 

 

 

 

 

 

 

 
 

 

 

 

 

 

 

 

 
 
 

 
 
 

 
 
 

 
 
 

 

Содержание BS-3520

Страница 1: ...BS 3520 Digital Indicator The Better Way for Weighing Measurements ...

Страница 2: ...eplace Fuse 12 6 Calibration 13 6 1 Calibration Mode 13 6 2 Digital Calibration Mode 14 6 3 Actual Load Calibration Mode 16 6 4 Linearization Calibration Mode 18 6 5 Digital Zero Calibration Mode 21 7 Function Mode 24 7 1 Function Setting Method 24 7 2 Function Items 25 8 HOLD 30 8 1 Hold Mode 30 8 2 Hold Type 32 8 3 Hold Operation 33 9 Relay Mode 35 9 1 Relay Mode 35 9 2 Relay Comparator Mode 38 ...

Страница 3: ...1 BCD Output 51 11 1 BCD Parallel Output 51 12 Serial Output 52 12 1 RS 232C 422 485 Serial Interface 52 12 2 Format 54 13 Check Mode 59 13 1 Operation for Each Check Mode 59 14 Key Lock Mode 62 14 1 Key Lock Method 62 14 2 Key Lock Cancellation Method 62 15 Repair 63 15 1 Error Message 63 15 2 Load Cell Inspection 63 15 3 Load Cell Access Diagnosis 64 15 4 Pattern of Display Letters 64 16 Warrant...

Страница 4: ...BS 3520 is high precision high speed indicator of 96x96mm size High precision 24bit Sigma Delta A D converter High speed A D and D A conversion of 2000 time sec A D external resolution 1 20 000 Actual load or digital calibration Linearity Compensation Function 4 point excluding zero External display of min max setting value Low OK and High relay contact output Hold or Peak hold function Basic equi...

Страница 5: ...witch before connecting peripherals Grounding of equipment shall be conducted in order to prevent electric noise and fall Exertion of voltage or current over maximum allowable value will lead to damage in the product Power voltage shall be set within allowable range Use outside allowable range may cause fire electric shock and defect Please understand the fact that contents of manual may be change...

Страница 6: ... 5digit Range of Max Display 19999 19999 min gradation 1 199950 999950 min gradation 50 Display Conversion Speed 1 000 times sec 1 time sec Polarity indication minus sign Annunciators LO OK HI HOLD ZERO Display increments 1 2 5 10 20 50 selectable Decimal Point Location 0 0 0 0 00 0 000 0 0000 selectable to any points Power Source AC 85 240V 50 60Hz Free Voltage Power Consumption Approx 20VA Range...

Страница 7: ...ze PANEL CUT 92 92 91 91 96 96 6 126 9 141 EXC EXC COM2 L 3 L 2 L 1 ZERO COM1 HOLD 6 INPUT OUTPUT LOAD CELL 7 8 9 10 11 12 1 2 3 4 5 CURR VOLT AC85 AC240V 48Hz 63Hz M 1 2 3 4 LO LOW HIGH OK HI HOLD ZERO BS DIGITAL INDICATOR 3520 ...

Страница 8: ...tion mode Status Lamp LO It flickers when measurement value is under min setting value It flickers when it is lower or higher than RY1 relay setting value OK It flickers when low value measurement value max value It flickers when it is lower or higher than RY2 relay setting value HI It flickers when measurement value exceeds max setting It flickers when it is lower or higher than RY3 relay setting...

Страница 9: ...o 4 key or No 4 key while pushing No 3 key It is used when entering function mode No 3 key while pushing No 1 key or No 1 key while pushing No 3 key It is used when entering calibration mode No 3 key while pushing No 2 key or No 2 key while pushing No 3 key It is used when entering relay setting mode No 2 key while pushing No 1 key or No 1 key while pushing No 2 key It is used when entering hold m...

Страница 10: ...AC240V 48Hz 63Hz EXC EXC COM2 L 3 L 2 L 1 ZERO COM1 HOLD 6 INPUT OUTPUT LOAD CELL 7 8 9 10 11 12 1 2 3 4 5 CURR VOLT AC85 AC240V 48Hz 63Hz Load Cell Input Output Option Grounding Power Serial Output Connection 5 Description on Rear Panel Standard Specification Analog Output Option BCD Output Option 5 1 Description on Each Terminal Unit 5 1 1 Power AC IN Terminal No Name Contents L AC IN Power Inpu...

Страница 11: ...shall be separated and then installed Please use shield cable for load cell cable In case of extending the load cell cable use of load cell cable of our company is recommended Shield cable of load cell cable shall be connected to BS 3520 terminal only In case of sealing road cell input cable SIG green and SIG blue cable shall be connected to No 4 and No 3 respectively There may be abnormality in r...

Страница 12: ... 1 4 Output 9 10 11 12 Terminal No Name Contents 9 L1 LO RY1 relay output terminal a b contact point output based on mode setting 10 L2 OK RY2 relay output terminal a b contact point output based on mode setting 11 L3 HI RY3 relay output terminal a b contact point output based on mode setting 12 COM2 Common terminal for relay output Output method differs based on relay mode setting Refer to relay ...

Страница 13: ...ry of cable to terminal may cause the damage of terminal Do not insert cable to top of terminal to insert driver 5 2 2 How to replace fuse 1 Loosen the bolt of rear panel and pull off PCB as illustrated in the figure 2 Replace the fuse Fuse Capacity 250V 1 0A 20mm 3 Assembly shall be conducted in reverse of disassembly Warning It shall be conducted only when the power is turned off EXC EXC COM2 L ...

Страница 14: ...ode It is calibration mode to conduct non linear calibration that reduces the measurement error through calibration of maximum 4 points excluding zero with the use of actual load 6 1 2 How to select calibration mode 1 It turns into the mode selection status when key is pushed while pushing key is pushed at measurement status 2 Mode change and setting value change is available when key or key is pu...

Страница 15: ... HOLD ZERO BS DIGITAL INDICATOR 3520 6 2 Digital calibration mode 6 2 1 Calibration Method Step 1 Decimal point setting and min gradation setting When you push No 3 key while pushing No 1 key entry to function mode is available It is changed to decimal point mode when you push No 4 key Push No 4 key by changing decimal point by pushing No 2 or No 3 key Input is completed when you push No 4 key aft...

Страница 16: ...1 2 3 4 Input load cell rated output using No 2 and No 3 key and Push No 4 key Ex 2 0000 mV V Input rated capacity of load cell with the use of No 2 and No 3 key ex 15 000 kg Digital input is completed when pushing No 4 key and it returns to measurement mode Step 3 Zero Calibration Push No 3 key while pushing No 4 key Zero calibration is completed Warning Although the setting range of resolution i...

Страница 17: ...GH OK HI HOLD ZERO BS DIGITAL INDICATOR 3520 6 3 Actual load calibration mode 6 3 1 Calibration Method Step 1 Decimal point setting and min gradation setting Entry to function mode is available when pushing No 3 key while pushing No 1 key It changes to decimal point mode when pushing No 4 key Push No 4 key after changing decimal point by pushing either No 2 or No 3 key Input is completed when you ...

Страница 18: ...ith the use of No 2 and No 3 key Ex 5 000kg Push No 4 key after exerting actual load to load cell ErrO is played and it is turned to LoAd status when load is small or direction of load is applied toward direction Actual load calibration has been completed Check the dead weight value indicated to display Check the zero point return status after lowering the dead weight and Step2 shall be executed r...

Страница 19: ... calibration mode which conducts non linear calibration that reduces the measurement error by conducting calibration with max 4 points excluding the zero with the use of actual load ZERO 0 Weight Value Display Error ZERO Before linear compensation After linear compensation 0 Weight Value Display Calibration point 1 Load1 Calibration point 2 Load2 Calibration point 3 Load3 SPAN SPAN Step 1 Decimal ...

Страница 20: ...CATOR 3520 Step 2 Zero calibration and dead weight setting Push No 2 key while pushing No 3 key It is changed into L CAL mode when pushing either No 2 or No 3 key Enter into L CAL mode by pushing No 4 key Check the stability after display of UnLoad and push No 4 key when there is nothing on load cell Conduct zero calibration Input dead weight value by using No 2 and No 3 key after entering 1st dea...

Страница 21: ...d eve when it exceeds 20000 Check whether or not regular measurement is conducted and carry out calibration upon the necessity Calibration under instable environment may cause the measurement error Load 1 Load 2 shall be input for Load 2 dead weight input value Load 1 Load 2 Load 3 for Load 3 dead weight input value is and Load 1 Load 2 Load 3 SPAN for SPAN dead weight input value In regards to lo...

Страница 22: ... is not conducted on the hold When offset value is set at function the value other than offset value shall be executed as zero Setting shall be conducted at function in order execute initial zero when turning on the power Zero calibration may not be conducted when there is mechanical interference or load cell abnormality 6 5 2 Zero Calibration Method by External Input It shall be executed when not...

Страница 23: ...nE None Width of zero tracking shall be set 0 000 Setting Range 0 999 Time of Zero Tracking 00 Setting Range 0 99 Zero tracking operation time shall be set 00 Operation within 0 1 sec 01 0 1 sec 99 9 9 sec Warning When None is selected at zero tracking width setting zero tracking time setting mode shall not be displayed Display 0 Tracking width Start Tracking time Zero calibration cannot be conduc...

Страница 24: ...bered even when the power is turned on Item Display Setting Value Setting Contents Automatic Zero Point Mode nonE None It is a function to automatically execute the zero calibration with power input upon Power On AUto Automatic Zero Point It cannot be returned when it is turned into zero point after automatic execution of zero point When offset value is set at function automatic zero point is exec...

Страница 25: ... pushed at measurement status 2 Mode change and setting value change is available when key or key is pushed at mode selection status 3 Push key in order to move onto next stage after saving the setting value at mode selection status 4 Push key to cancel the setting and return to measurement mode for mode cancellation At the mode cancelation the setting until previous stage of cancellation is saved...

Страница 26: ... 4 digit Min Gradation 1 0 1 2 3 4 Min gradation shall be set Warning Random change is available even after the completion of calibration However the calibration shall be conducted again in case of decimal point change change in number of digits 2 0 2 4 6 8 5 0 5 10 15 20 10 0 10 20 30 40 20 0 20 40 60 80 50 0 50 100 150 Display Conversion Speed 1 4 7 times sec Slow It is display conversion speed ...

Страница 27: ...old and external contact point NonE None EdGE Edge Hold PK PK Hold Relay Relay Mode Select relay mode Warning In case it is set as None nothing is displayed to low and high display unit bASIC Basic limit mode HY_HI High limit mode HY_Lo Low limit mode HY_LLH Low Low High limit mode NonE None Hysteresis Width 00 000 Setting Range 0 999 Width of hysteresis shall be set Zero Calibration Zero Tracking...

Страница 28: ... 20mA output 0 20mA 0 20mA output 5V 0 5V output Analog Lo Output 000 000 Setting Range 199999 999999 In case analog output is 0V 0mA 4mA display value shall be set Analog output is conducted by force while digit is flickering in case only one digit is flickering Analog High Output 010 000 Setting Range 199999 999999 In case analog output is 5V 10V 20mA display value shall be set Analog output is ...

Страница 29: ...n ID 1 Setting Range 00 99 Communication address shall be selected Communication Speed Communication transmission speed shall be set 1200 1200 bps 2400 2400 bps 4800 4800 bps 9600 9600 bps 19200 19200 bps 38400 38400 bps 57600 57600 bps 115200 115200 bps Data Length Communication data length shall be set Warning For 7bits setting the parity bit shall be set as either even or odd When it is set as ...

Страница 30: ...ata request Refer to communication format Page 54 56 Communication format Select communication format Warning In case of AND Format setting command mode is same as the 12 2 2 command mode bS3520 BONG SHIN Format And AND Format Communication RS 422 485 setting Select RS 422 485 communication Warning After changing from RS 485 to RS 422 turn the power off and then use is after turning it on Please u...

Страница 31: ...Method by external hold input terminal Hold is commenced when external hold input terminal is turned on contact point input At the moment external hold input terminal shall maintain the on status Hold is cancelled when input is conducted at off status Method by serial output RS 232C 422 485 command order Hold is commenced by the hold on command with communication and hold is cancelled by hold off ...

Страница 32: ...t mode when pushing No 1 key Function setting has been completed and setting value shall be saved It is changed to measurement mode status Entry to setting mode is unavailable under key lock status thus it shall be executed after canceling the key lock Setting range display may differ based on location of decimal point In case of setting hold function as None key hold and external contact point ho...

Страница 33: ...ent of hold input Load cell input Display value ON Front hold key OFF Hold section ON External hold input OFF 8 2 2 Peak hold mode PK hold conducts hold of PK display and output at the moment of hold input Load cell input Display input ON Front hold key OFF Hold section ON External hold input OFF ...

Страница 34: ...ced Load is exerted in case of PK hold Display is made with PK hold In case of edge hold setting hold is conducted for display and output at the moment of key hold input In case of PK hold setting holding of PK display and output is conducted before cancelation of hold 8 3 2 Operation Cancelation by Hold Key Hold OFF Hold is cancelled when pushing No 1 key while pushing No 4 key Hold is cancelled ...

Страница 35: ...conducted when hold mode is not set as edge or PK hold from function Switch shall be zero voltage input of machine contact point and semiconductor contact point Zero calibration is not conducted under Hold status ON OFF time shall be secured for 10ms or longer Please use it within rated load 8 3 4 Method by serial output RS 232C 422 485 command Hold is commenced by hold on command through communic...

Страница 36: ...atus when key is pushed while pushing key is pushed at measurement status 2 Relay mode selection item comes out when you push key 6 times at mode selection status 3 Mode change is available when key or key is pushed at mode selection status 4 Push key in order to move onto next stage after selecting the relay mode and saving 5 Push key to cancel the mode and return to measurement mode It is entere...

Страница 37: ...ting value change mode when pushing No 2 key while pushing No 1 key When pushing No 4 key after changing Lo relay setting value with the use of No 2 and No 3 key it moves onto next stage ex 1 000kg When pushing No 4 key after changing Hi relay setting value with the use of No 2 and No 3 key it turns into measurement mode ex 3 000kg Relay setting has been completed and setting value is saved It tur...

Страница 38: ... It is entered into relay setting value change mode when pushing No 2 key while pushing No 1 key When pushing No 4 key after changing RY1 relay setting value with the use of No 2 key and No 3 key it moves onto next stage ex 1 000kg When pushing No 4 key after changing RY2 relay setting value with the use of No 2 key and No 3 key it moves onto next stage ex 2 000kg When pushing No 4 key after chang...

Страница 39: ... Hi setting value High limit mode Relay output terminal Output condition L1 LO Measurement value RY1 setting value L2 OK Measurement value RY2 setting value L3 HI Measurement value RY3 setting value Low limit mode Relay output terminal Output condition L1 LO Measurement value RY1 setting value L2 OK Measurement value RY2 setting value L3 HI Measurement value RY3 setting value Low Low High limit mo...

Страница 40: ...tting value RY2 setting value 9 2 2 Example of relay operation Basic limit mode Relay output terminal Output condition L1 LO Measurement value Lo setting value L2 OK Lo setting value Measurement value Hi setting value L3 HI Measurement value Hi setting value High limit mode Relay output terminal Output condition L1 LO Measurement value RY1 setting value L2 OK Measurement value RY2 setting value L3...

Страница 41: ...N OFF RY1 setting value RY3 setting value RY2 setting value Low limit mode Relay output terminal Output condition L1 LO Measurement value RY1 setting value L2 OK Measurement value RY2 setting value L3 HI Measurement value RY3 setting value Low Low High limit mode Relay output terminal Output condition L1 LO Measurement value RY1 setting value L2 OK Measurement value RY2 setting value L3 HI Measure...

Страница 42: ...set from function 1 Basic limit mode Relation between LO and OK When the measurement value becomes higher than Lo setting value it immediately turns into OK When the measurement value is lower than Lo value at OK status it does not immediately become Lo but it turns into Lo when it became lower than width of hysteresis Relation between OK and Hi When the measurement value becomes higher than Hi se...

Страница 43: ... the measurement value becomes higher than RY2 OK setting value When the measurement value becomes lower than setting value under On status of RY2 OK it is not immediately turned off when the measurement value becomes lower than setting value but it is turned off when it becomes lower than width of hysteresis RY3 HI output It is immediately turned on when the measurement value becomes higher than ...

Страница 44: ...asurement value becomes higher than setting value under On status of RY2 OK it is not immediately turned off but it is turned off when it becomes higher than width of hysteresis It is immediately turned on when the measurement value becomes lower than RY2 OK setting value RY3 HI output When the measurement value becomes higher than setting value under On status of RY3 HI it is not immediately turn...

Страница 45: ... measurement value becomes lower than RY1 LO setting value RY2 OK output When the measurement value becomes higher than setting value under On status of RY2 OK it is not immediately turned off but it is turned off when it becomes higher than width of hysteresis It is immediately turned on when the measurement value becomes lower than RY2 OK setting value RY3 HI output It is immediately turned on w...

Страница 46: ...Output Mode 10 1 1 Analog output selection mode 1 It turns into the mode selection status when key is pushed while pushing key is pushed at measurement status 2 Relay mode selection item comes out when you push key 11 times at mode selection status 3 Mode change is available when key or key is pushed at mode selection status 4 Push key in order to move onto next stage after selecting the analog mo...

Страница 47: ...100ppm C typ 100ppm C typ Non linearity Within 0 05 F S Within 0 05 F S Load resistance 5 kΩ or higher 500Ω or higher 10 2 2 Connection method DAV Terminal No Contents VOLT Voltage output VOLT Voltage output 10 2 3 Connection method DAI Terminal No Contents CURR Current output CURR Current output Analog Output is to convert measurement value into D A for output and display conversion speed shall b...

Страница 48: ...en key or key is pushed at mode selection status 4 Push key in order to move onto next stage after selecting the analog output setting and saving 5 Push key to cancel the mode and return to measurement mode It is entered into function mode when pushing No 3 key while pushing No 1 key It moves onto analog Lo output setting stage when pushing No 4 key 12 times It moves onto next stage when pushing N...

Страница 49: ...10V 0V DAC LO DAC HI 5V 10V 5V 10 3 2 Analog output graph DAV In case of setting DAC LO and DAC HI as same value output does not increase from 0V Reverse output is only applicable at 0 10 5V setting It is not applicable at 0 5V 0 10V 0 20mA 4 20mA output For reverse output in case of setting DA Lo setting value and DAC Hi setting value is each set as 10 000 and 0 000 0V and 10V 5V is each output f...

Страница 50: ... in performance when meter with low precision is used 10 4 1 Analog output zero span minute calibration 1 It turns into the mode selection status when key is pushed while pushing key is pushed at measurement status 2 Relay mode selection item comes out when you push key 14 times at mode selection status 3 Mode change is available when key or key is pushed at mode selection status 4 Push key in ord...

Страница 51: ...9999 4 0 mA increase decrease 0 20mA output 0 9999 9999 0 0 mA 10 10 0 003 mA increase decrease 20 20 0 006 mA increase decrease 100 100 0 03 mA increase decrease 999 999 0 3 mA increase decrease 9999 9999 3 0 mA increase decrease Calibration range of voltage output is max 4 0V and current output is 4 0mA Entry to setting mode is unavailable under Key Lock status thus it shall be executed after ke...

Страница 52: ...n 20 Output 4 10 4 2 Output 1 10 0 21 Output 8 3 Output 2 22 Output 1 10 5 4 Output 4 23 Output 2 5 Output 8 24 Output 4 6 Output 1 10 1 25 Output 8 7 Output 2 26 Output Polarity 8 Output 4 27 Output Decimal point 10 1 9 Output 8 28 Output Decimal point 10 2 10 Output 1 10 2 29 Output Decimal point 10 3 11 Output 2 30 Output Decimal point 10 4 12 Output 4 31 Output Decimal point 10 5 13 Output 8 3...

Страница 53: ... 8 bits Parity bit None Even Odd Stop bit 1bit Use code ASCII Terminator STX ETX STX 02H ETX 03H No of max connection 1 Max 32sets RS 422 10sets Connector specification D SUB 9 pin Female male for cable side 12 1 2 Connection method For RS 232C For RS 422 Warning In case of using RS 422 change shall be made from function setting For RS 485 Pin No Signal name Direction Description 2 TXD Output Tran...

Страница 54: ...48Hz 63Hz BS 3520 6 AC85 AC240V 48Hz 63Hz BS 3520 6 7 7 12 1 3 Example of RS 485 connection 1 1 1 connection with master Shield cable and twist pair cable shall be used Longitudinal resistance shall be exerted to both ends of writing 100Ω 120Ω 1 2W 2W There may be a reverse display of A and B for some master equipment Connect by changing A and B in case communication is unavailable 2 Multiple conn...

Страница 55: ...5 6 L ETX STX 0 1 1 2 3 4 5 6 O ETX STX 0 1 1 2 3 4 5 6 H ETX STX 0 1 1 2 3 4 5 6 H ETX Warning When there is no decimal point in display decimal point is received at the end of weight data Response example Hexa ID Sign Weight data Decision 1 2 3 4 5 6 7 8 9 10 11 12 13 02 30 31 2B 31 32 33 2E 34 35 36 4C 03 02 30 31 2B 31 32 33 2E 34 35 36 4F 03 02 30 31 2B 31 32 33 2E 34 35 36 48 03 02 30 31 2B ...

Страница 56: ...le ID Command 1 2 3 4 5 STX 0 1 Z ETX 02 30 31 5A 03 3 Hold ON command Hold is commenced Command example ID Command 1 2 3 4 5 STX 0 1 H ETX 02 30 31 48 03 4 Hold OFF command Hold is cancelled Command example ID Command 1 2 3 4 5 STX 0 1 C ETX 02 30 31 43 03 ...

Страница 57: ...0 0 0 ETX STX 0 1 R Y 3 0 3 0 0 0 ETX Data reception command and response toward command differ based on ID setting and location of decimal point Serial output and command setting can be conducted even under hold status only zero command cannot be operated Command setting is available even under stream mode setting operation of zero command is also available However since data may be broken due to...

Страница 58: ...Response example with decimal point ID COMMAND Sign VALUE 1 2 3 4 5 6 7 8 9 10 11 12 13 14 STX 0 1 R L O 0 1 0 0 0 ETX STX 0 1 R H I 0 3 0 0 0 ETX STX 0 1 R Y 1 0 1 0 0 0 ETX STX 0 1 R Y 2 0 2 0 0 0 ETX STX 0 1 R Y 3 0 3 0 0 0 ETX Response example without decimal point Number of digit is reduced by one ID COMMAND 부호 VALUE 1 2 3 4 5 6 7 8 9 10 11 12 13 STX 0 1 R L O 0 1 0 0 0 ETX STX 0 1 R H I 0 3 ...

Страница 59: ...g Header 1 Header 2 Data Unit Terminator 8 digit with a polarity and a decimal point 12 2 4 ASCII Code Lower bits 0 1 2 3 4 5 6 7 Upper bits 0 Space 0 P Space p 1 1 A Q a q 2 2 B R b r 3 3 C S c s 4 4 D T d t 5 5 E U e u 6 6 F V f v 7 7 G W g w 8 8 H X h x 9 9 I Y i y A LF J Z j z B K k C L l D CR M M E N N F O _ o ASCII code Hexadecimal Description Header 1 ST 53 54 Stable UN 55 4E Unstable OL 4F...

Страница 60: ...ation At the mode cancelation the setting until previous stage of cancellation is saved It is entered into mode when turning the power on while pushing No 2 and No 3 key under power off status Current load cell output voltage is displayed 000 000mV is displayed under no load and mV is differently displayed based on dead load Load cell shall be examined with the display of higher than 015 000mV or ...

Страница 61: ...ion 13 1 4 Analog output DAC voltage current check It is a mode to check analog output voltage current Analog high output setting value is displayed to meter display unit Check output voltage current with the equipment of digital multi meter to the outside When there is change in analog high output setting value output of voltage current is conducted based on display value It is a mode to check an...

Страница 62: ...ing No 4 key and push No 1 key for cancellation Check mode is displayed only when BCD is selected from 13 1 3 Optional output DAC BCD selection mode For range of setting value change setting is available within 199999 999999 Decimal point and output are also available initial value is set as 000 000 13 1 6 Initialization Mode Various settings of data are initialized All settings including calibrat...

Страница 63: ... DIGITAL INDICATOR 3520 M 1 2 3 4 LO LOW HIGH OK HI HOLD ZERO BS DIGITAL INDICATOR 3520 M 1 2 3 4 14 Key Lock Mode It is a key lock mode It is a mode to protect calibration zero setting and function item setting value Entry to relay setting value change mode is available only 14 1 Key lock method Push No 2 key while pushing No 4 key LoCK is displayed and it is turned into key lock mode 14 2 Key lo...

Страница 64: ...oad cell shall be checked since load cell output is 35mV or lower It exceeded max range of display value It exceeded max range of display that is either 199999 or 999999 Conduct calibration again after checking load cell output voltage and disconnection 15 2 Load cell inspection When the load cell in use is instable following items shall be inspected 1 Check whether or not load cell terminal is pr...

Страница 65: ...o 13 1 2 load cell output voltage check It is part measured when checking the connection of load cell In case of using summing box same measurement shall be conducted inside the summing box as well Diagnosis item Diagnosis location Criteria normal standard 1 Load cell applied voltage EXC EXC DC 5V 2 Load cell output voltage SIG SIG Within 15mV 3 Load cell focused voltage EXC SIG Around DC 2 5V EXC...

Страница 66: ...rmined to be ours we provide A S free of charge Forwarding of this product is conducted through strict quality management and inspection process Please contact the agency or our company in case of defect occurrence A S and Product Inquiry TEL 82 31 742 6661 FAX 82 31 742 6664 www bongshin com e mail loadcell bongshin com www loadcell co kr ...

Отзывы: