10-2 Analog output specification and connection method (DAV, DAI)
For DAV(DAI) output, the measurement value is converted into D/A and output as analog voltage (current).
Output range include 0~5V, 0~10V, 0 ~ ±5V, 0 ~ ±10V, 0~20mA, and 4~20mA.
Output scaling and minute calibration is set from function.
Frequency trait is based on display conversion speed setting of function.
10-2-1 Specification
Range of rated output
–10 V ~ +10 V
4 ~ 20mA
Range of max output :
–12.0 V ~ +12.0 V
0 ~ 24mA
D/A conversion speed :
1000
times/sec
1000 times/sec
Valid resolution :
1/10000 1/10000
Zero drift :
±1 mV/°C typ.
±0.5uA/°C typ.
Gain drift :
±100ppm/°C typ.
±100ppm/°C typ.
Non-linearity :
Within ±0.05%F.S.
Within ±0.05%F.S.
Load resistance ::
5 k
Ω or higher
500
Ω or higher
10-2-2 Connection method
(DAV)
Terminal No.
Contents
VOLT +
Voltage
VOLT -
Voltage output -
10-2-3 Connection method
(DAI)
Terminal No.
Contents
CURR +
Current
CURR -
Current output -
Analog Output is to convert measurement value into D/A for output and display conversion speed shall be
changed from function setting in case of wanting 1000 times/sec output.
It is output based on display conversion speed of function setting.
In
case
of
DAI Analog Output, (-) display value is not output.
Output
at
function
mode
entry
maintains
previous
value
before
entry
and
there
is
no
output
change.
Voltage
and
current
output
cannot
be
used
at
once.
Either
voltage
or
current
mode
shall
be
selected
from
function
setting
for
use.
(it
is
different
output
terminal).