10-4 Analog output zero and span minute calibration
Minute calibration is conducted for measurement value equivalent to 0V(0mA, 4mA), and 10V(20mA)
of DAV (DAI) output.
Output minute calibration is set from function.
This mode is not displayed with the equipment of BCD OPTION.
Warning
This
mode
shall
be
used
only
when
there
is
lack
of
accuracy
even
after
analog
output
zero
and
span
calibration.
High
precision
digital
multi
‐
meter
is
necessary
in
order
to
conduct
calibration
since
it
controls
inner
calibration
factor.
There
may
be
deterioration
in
performance
when
meter
with
low
precision
is
used.
10-4-1 Analog output zero & span minute calibration
1.
It turns into the mode selection status when key is pushed while pushing key is
pushed at measurement status.
2.
Relay mode selection item comes out when you push key 14 times at mode selection status.
3.
Mode change is available when key or key is pushed at mode selection status.
4.
Push key in order to move onto next stage after selecting the analog output setting value and saving.
5.
Push key to cancel the mode and return to measurement mode.
It is entered into function
mode when pushing No.3
key while pushing No.1 key.
It moves onto stage for
minute calibration of analog
zero output value when
pushing No.4 key 14 times.
It moves onto next stage
when pushing No.4 key after
changing zero output value
and saving with the use of
No.2 and No.3 key.
It moves onto to next stage
when pushing No.4 key when
changing span output value
with the use of No.2 and No.3
key.
It is changed into
measurement mode when
pushing No.1 key upon
setting completion.
Function setting has been
completed and it is changed
into measurement mode
status after saving the setting
value.