21.
Command
: STATus:OPERation:ENABle?
Description :
Read Status Operation Enable Register Query.
Function :
This query command reads the contents of the Status Op-
eration Enable Register, in the <NR.DATA> (0-32767) for-
mat, whose corresponding HEX value represents the bit pat-
tern of the Register.
22.
Command
: STATus:OPERation:CONDition?
Description :
Read Status Operation condition register query.
Function :
This query will give the contents of Status Operation Condition
Register in <NR.DATA> (0-32767) format, whose HEX value
represents the Bit pattern of the register. This will always
read as ‘0’ (ZERO).
23.
Command
: STATus:PRESet
Description :
Resets the Enable Bits of Status Operation and Question-
able Enable Register.
Function :
This command puts the data as ‘0’ (ZERO) so that corre-
sponding HEX data (00 Hex) will CLEAR the Status Opera-
tion Enable and Status Questionable Enable Registers. (to
disable all the corresponding events in the Status Opera-
tion Event Register and Status Questionable Data Event
Register ).
6.7.1
*RST Conditions
The following conditions are defined for the PROGRAMMABLE DC POWER
SUPPLY on receiving the *RST command and execute it.
1).
The Baud Rate of the instrument for RS232 operation is set to 9600 BAUD,
which is variable through Remote mode commands.
2).
The output Voltage, Current, OVP Limit and OCP Limit for all the Channels are
set to the same values as programmed in the Configuration setup, stored in the
EEPROM installed inside the unit.
3).
The output ON/OFF relays for all the Programmable Channels are switched to
‘OFF’ position.
6.8
INSTRUMENT (SYSTEM RELATED) COMMANDS
This section explains in detail the Instrument Specific or the System Related
Commands, which are used to Program different parameters of the system. All
commands, except the last two commands (For selecting the Internal or External
SENSE for the output) are SCPI APPROVED commands. But the last two are SCPI
CONFIRMED commands. For this instrument there is NO command, which are not
part of the SCPI definition.
51
Содержание VSP12010
Страница 1: ...MODELS VSP6020 VSP2050 VSP4030 VSP12010 High Power Switching DC Power Supplies Instruction Manual...
Страница 14: ...12...