BIT DEFINITIONS-QUESTIONABLE DATA REGISTER
BIT
DESCRIPTION
DECIMAL
VALUE
DEFINITION
0
OVP trip
1
Channel 1 Over Voltage Trip
2
OT trip
4
Channel 1 Over Temperature Trip
3
OVP trip
8
Channel 2 Over Voltage Trip
5
OT trip
32
Channel 2 Over Temperature Trip
6
Not Used
64
Always Set to 0
7
Not Used
128
Always Set to 0
8
OVP trip
256
Channel 3 Over Voltage Trip
10
OT trip
1024
Channel 3 Over Temperature Trip
11
OVP trip
2048
Channel 4 Over Voltage Trip
13
OT trip
8192
Channel 4 Over Temperature Trip
14
Not Used
16384
Always Set to 0
15
Not Used
32768
Always Set to 0
The questionable data event register is cleared when:
* You execute a *CLS (clear status) command.
* You query the event register using STATus: QUEStionable: EVENt?
The questionable data enable register is cleared when:
* You turn on the power to the instrument.
* You execute the STATus: PRESet command
* You execute the STATus: QUEStionable: ENABle 0 command.
6.6
COMMAND SUMMARY
This section summarizes the SCPI commands available for this Programmable
Instrument. Refer to the later sections in this chapter for more complete details on each
command.
Throughout this manual, the following conventions are used for SCPI command
syntax. Square brackets ([ ]) indicate optional keywords or parameters. Braces ({})
enclose parameters within a command string. Triangle brackets (< >) indicate that
you must substitute a value for the enclosed parameter.
44
Содержание VSP12010
Страница 1: ...MODELS VSP6020 VSP2050 VSP4030 VSP12010 High Power Switching DC Power Supplies Instruction Manual...
Страница 14: ...12...
Страница 42: ...SCPI STATUS SYSTEM 40...