Command: VOLT: PROT<SP><NR.DATA><EOS>
Description: Set Over Voltage Protection Limit of OUTPUT
Function: This command sets the Over Voltage Protection Limit of the Power Supply
output with the <NR.DATA> given in the command in Volts. The input VOLTAGE data
is in Numeric Data format. Once this OVP limit is set, any time if the actual Output
Voltage of the channel exceeds the OVP limit, the OUTPUT will be ‘TRIPed’. I.e. the
OVP operates to protect the output and the O/P voltage will be made zero and the
OUTPUT ON/OFF relay will be switched OFF.
For example to set OVP of 31.50 V for output, send the command as
VOLT: PROT 31.50<EOS>
5.
Command
: OUTP<SP>ON<EOS>
Description: Turn ON the output of power supply.
Function: This command is used to make the OUTPUT available at the output
terminals.
6.
Command
: OUTP<SP>OFF<EOS>
Description: Turn OFF the output of power supply.
Function: This command is used to switch OFF the output & thus to make the
OUTPUT zero at the output terminals.
7.
Command
: SYST: REM<EOS>
Description: Takes the instrument into Remote Mode of operation
Function: This command will explicitly take the unit to remote mode. All the front
panel controls are disabled.
8.
Command
: SYST: LOC<EOS>
Description: Take the instrument to Local mode of Operation.
Function: This command will take the unit to Local mode of operation & ALL the front
panel controls are enabled.