BT-720 Operation manual
P/N:720-ENG-OPM-ENG-R03
Bistos Co., Ltd.
2020.10
7
this manual.
•
Do not remove the covers of a BT-720 yourself to avoid damage to the equipment
and unexpected electrical shock. Only qualified Bistos service engineer must repair
or replace components.
CAUTION
•
Please install or carry the instrument properly to prevent damage due to falling,
collision, strong vibration or other mechanical force.
•
Avoid instrument splashed by water.
•
Avoid high temperatures, the instrument should be used within a temperature
range of 5
℃
~ 40
℃
。
•
Avoid using instrument in the environment such as pressure is too high, poor
ventilation, dusty, or contain salt, sulfur gas and chemical.
•
Before using the monitor, check the monitor and accessories if there is damage
that may affect patient safety. If there is obvious damage or aging, replace the
parts before use. The replacement should be made with same parts of original
parts.
•
Before powering on the device, make sure that the power used by the device
complies with the supply voltage and frequency requirements on the equipment
label or in the Operator’s Manual.
•
Equipment should be tested at least once a year, the test should be done and
recorded by trained, have security testing knowledge and experienced personnel.
If there are any problems in the tests, they must be repaired.
•
When the instrument and accessories are about to exceed the useful life
(expected service life: 5 years), it must be treated in accordance with relevant
local laws and regulations or the hospital's rules and regulations.
•
Do not connect to other equipment or network which not specified in the
instruction for use, in risk of external high voltage.
•
Do not connect any equipment or accessories that are not approved by the
manufacturer or according to IEC 60601-1 to the monitor. The operation or use of
non-approved equipment or accessories with the monitor is not tested or
supported, and monitor operation and safety are not guaranteed in such a case.
•
Any non-medical equipment (such as the external printer) is not allowed to be
used within the patient vicinity (1.5m/6ft.).
•
Parts and accessories used must meet the requirements of the applicable safety
standards, and/or the system configuration must meet the requirements of the
medical electrical systems standard.
•
Ensure that the conductive parts of electrodes and associated connectors,
including neutral electrodes, do not come in contact with earth or any other
conducting objects.
•
Protection of ME EQUIPMENT against effects of discharge of a cardiac defibrillator
depends on use of proper cables.
0.2 Shock hazards
WARNING
Unplug the monitor from its power source prior to cleaning or maintenance to prevent
personal injury or equipment damage.
Some chemical cleaning agents may be conductive and leave a residue that may permit
a build-up of conductive dust or dirt. Do not allow cleaning agents to contact electrical