BT-720 Operation manual
P/N:720-ENG-OPM-ENG-R03
Bistos Co., Ltd.
2020.10
24
1.
Prepare enzyme cleaning agent, distilled water and 10% solvent, respectively in different spray
bottle.
2.
Sprinkle cleaning agent on NIBP cuff, connector and extension tube, keep 1 minute for the dry
stains.
3.
Use a soft cloth to wipe smooth face. Use soft hair brush to brush visible stain and irregular
surface
4.
Rinsed with copious amounts of distilled water.
NOTE
Please be especially careful to clean the air ball and control valve of whole air system.
Do not allow any liquid entering into reversing valve and saturated valve.
Don’t use a soft cotton ball and fiber to clean this accessory because they will stick on
the cuff and extension tube.
6.6.2 Disinfection method
1.
Sprinkle bleach solution (Formula: the proportion of water and bleaching powder to 1:10) then
keep 5 minutes
2.
Wipe off excess bleach solution and elute with distilled water again
3.
Natural dry cuff
7 Review
Use the Trend screen to recall all the historical patient data in a list, including monitoring time (in 1-
minute intervals), and SpO
2
, PR, SYS, DIA and MAP values. The most recent measurements display at
the top of the list.
Figure 7-1: SpO2 trend (1)
To view historical data:
Press the
[Setting] key twice to display the Trend screen.
Select
or
key to page up and down to view patient data.
Press key to exit the Trend screen and display the Main screen or press the [Setting] key
to see the trend graph.
Figure 7-2: SpO2 trend (2)