BT-720 Operation manual
P/N:720-ENG-OPM-ENG-R03
Bistos Co., Ltd.
2020.10
23
Select “Start Measure”, starts the blood pressure measurement; select “Stop Measure”, stop
the blood pressure measurement.
Select “Continue Measure”, continued measure pressure.
Select “Reset”, and restore the inflation pressure of the blood pressure pump to currently
configured initial settings. When the blood pressure pump is not working properly, but no
warning is given, you can reset the blood pressure pump, and automatically restores the blood
pressure pump.
Select
key to display the following interface.
Select “SYS Low Limit”, and enter value (Adult Range: 30-279, Pediatric Range: 30-229, Neonate
Rang: 30-144). Adult/ Pediatric /Neonate default: 90/70/40.
Select “SYS High Limit”, and enter value (Adult Range: 31-280, Pediatric Range: 31-230, Neonate
Rang: 30-145). Adult/ Pediatric /Neonate default: 160/120/90.
Select “DIA Low Limit”, and enter value (Adult Range: 10-219, Pediatric Range: 10-164, Neonate
Rang: 10-104). Adult/ Pediatric /Neonate default: 50/40/20.
Select “DIA High Limit”, and enter value (Adult Range: 11-220, Pediatric Range: 11-165, Neonate
Rang: 11-105). Adult/ Pediatric /Neonate default: 90/70/60..
Select “Alarm Level”, you can choose “Mid” or “High”.
Select “Default”, the alarm parameter is set to the default value.
Select “Calibration”, Users can not calibrate NIBP. If calibration is required, please contact your
service representative. Cuff pressure sensor should be checked and calibrated at least once
every two years by qualified professional service personnel.
Select “Leak Test”, the purpose of leakage test is to detect if the sealing of the air passage is in
good condition. If the leakage test passes, the alarm area displays “Leakage test Stopped”. If not
passed, the alarm area displays “Cuff leak” message. NIBP leakage test shall be at least once
every two years or when you think that the reading is not accurate.
6.6 Clean and disinfection method of NIBP cuff
If necessary, NIBP cuff and NIBP extension tube can be cleaned and disinfected together without
separated
6.6.1 Cleaning method