BT-720 Operation manual
P/N:720-ENG-OPM-ENG-R03
Bistos Co., Ltd.
2020.10
14
2.1.2 Placement requirements
Equipment installation must meet:
-
Ensure that the operating floor and the monitor have enough space for connecting the
accessory wires.
2.1.3 Power requirements
-
DC power supply adapter
Input: A.C. 100 V ~ 240 V, 50/60 Hz
Output: D.C. 15 V, 2.5 A
-
Built-in rechargeable lithium-ion battery: D.C. 11.1 V, 4400 mAh
2.1.4 Environmental requirements
The storage, transport and use of the monitor must meet the following environmental requirements.
Operating
environment
Ambient temperature
5
℃
~ 40
℃
Relative humidity
30 % ~ 85 % (Non-condensing)
Atmospheric pressure
700 ~ 1060 mbar (hPa)
Transportation
Prevent severe shock, vibration, rain and snow splashing during transport.
Storage
The packaged monitor should be stored in well-ventilated room with
ambient temperature -20
℃
~ 60
℃
, relative humidity 0 ~ 95 % (Non-
condensing), atmospheric pressure 700 ~ 1060 mbar(hPa), and without
corrosive gases.
The operating environment of the monitor should avoid noise, vibration, dust, corrosive or flammable
and explosive materials. In order to allow air flowing smoothly and achieve good heat dissipation, at
least 2 inches (5cm) clearance should be kept around the device.
When the device is moved from one environment to another, the device may have condensation due
to the differences in temperature or humidity. In this case, wait until the condensation disappears
before using the device.
WARNING
Ensure that the monitor is used under specified environment. Fail to do this, the
technical specifications declared in this manual may not be met and it may result in
damage to equipment and other unforeseen consequences.
2.2 Connecting to power
WARNING
Do not try to open the monitor when the power is connecting.
During the operation, do not disconnect any cable.
Connect to power adapter in the following steps:
-
Make sure that the AC power supply meets the following specifications: a.c.100V-240V,
50/60Hz.
-
Use the power adapter provided with the monitor. Plug the power adapter into the
power connector of the monitor, and plug the other end of the power adapter into the
mains (low voltage power supply network facilities) power outlet.
3 Basic operations
3.1 Turn on
3.1.1 Check the monitor
-
Before turn on the monitor, check whether there is mechanical damage to the monitor, and
whether the external cables and accessories are connected correctly.
-
Plug the power adapter into the AC power outlet. If using battery power, make sure the battery
is fully charged.
-
Check all the functions required for patient monitoring to make sure that the monitor operates
properly.