BT-720 Operation manual
P/N:720-ENG-OPM-ENG-R03
Bistos Co., Ltd.
2020.10
6
WARNING
•
Thoroughly read and understand the manual prior to use of the BT-720. Failure to
do so could result in personal injury or equipment damage.
•
The device is intended for clinical patient monitoring, and only trained and
qualified doctors and nurses should use the device.
•
The alarm volume, upper and lower alarm limits should be set according to the
actual situation of the using environment. Do not just rely on audio alarm system
while monitoring the patient, because too low alarm volume or muted alarm may
result in notice failure of alarm situation and endanger the patient’s safety. Please
pay close attention to the actual clinical status of the patient.
•
Use only the power adapter supplied with monitor.
•
Position the monitor where it is easy to de-energize the monitor when needed.
•
Do not open the enclosure to avoid an electric shock. Any repair and upgrade of
monitor should be done by service personnel trained and authorized by Bistos. Co.,
Ltd.
•
When handling packaging materials, abide by local laws and regulations or hospital
waste disposal regulations. Keep the packaging materials away from children.
•
Do not use in the presence of flammable anesthetics to prevent explosion or fire.
•
Install the power lines and cables of accessories carefully to avoid patient
entanglement or suffocation, cables tangled or electrical interference.
•
When the monitor is used together with electrosurgical devices, the user (a doctor
or a nurse) should ensure the safety of the patient and instrument.
•
The physiological wave, physiological parameters and alarm information displayed
on the monitor are only for the doctor’s reference and should not be directly used
as the basis for clinical treatment.
•
This is not a therapeutic device.
•
For patients with pacemakers, the cardio tachometer may count the pacemaker
pulse in case of a cardiac arrest or arrhythmias. Never rely solely on the cardio
tachometer alarm. Closely monitor the patients with pacemaker. For the inhibition
of the device on pacemaker, refers to this manual.
•
Use of accessories other than those listed and approved for use with this product
may result in increased emissions or decreased immunity.
•
Medical electrical equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in
this manual. In addition, portable and mobile RF communications equipment can
affect medical electrical equipment.
•
The equipment shall not be used adjacent to other devices unless verification of
normal operation in the configuration in which it is to be used can be achieved.
•
Keep matches, and all other sources of ignition, out of the room in which the
patient monitor is located. Textiles, oils, and other combustibles are easily ignited
and burn with great intensity in air enriched with oxygen. Personal injury or
equipment damage could occur.
•
A fire and explosion hazard exists when performing cleaning or maintenance
procedures in an oxygen-enriched environment.
•
The patient monitor has been validated with the accessories and options listed in
this manual and found to comply with all relevant safety and performance
requirements applicable to the device. It is therefore the responsibility of the
person or organization who makes an unauthorized modification, or incorporates
an unapproved attachment to the device.
•
An operator may only perform maintenance procedures specifically described in