BT-720 Operation manual
P/N:720-ENG-OPM-ENG-R03
Bistos Co., Ltd.
2020.10
28
WARNING
•
Before cleaning the monitor or sensor, turn off the power and disconnect the DC power.
•
The monitor should be kept clean. It is recommended to regularly clean the enclosure
surface and the display screen. Cleaning the enclosure with non-etching cleaner such as
soap and water.
CAUTION
•
To avoid damaging the monitor:
Do not use strong solvents such as acetone.
Most cleaners must be diluted before use. Diluting should be according to the
manufacturer's instructions.
Do not use abrasive materials (such as steel wool).
Do not allow any liquid entering into the enclosure, and never immerse any part of
the device into liquid.
Do not leave any cleaning solution on the surface of any part of the device.
NOTE
Wipe the monitor and sensor surface with medical alcohol, dry it naturally or with clean,
dry, lint-free cloth.
Bistos is not liable for effectiveness of using these chemicals for infectious disease
control. Please consult the infectious disease control officers or experts of the hospital
for advice.
10.3 Disinfection
In order to avoid damage to the product, we recommend that the product is disinfected only when it is
deemed necessary by the hospital maintenance procedures. We also recommend that the instrument
to be disinfected must first be cleaned.
CAUTION
•
To prevent damage to the monitor, do not disinfect the monitor with gas (EtO) or
formaldehyde.
11 Maintenance
WARNING
•
If the hospitals or institutions using this instrument can’t implement a satisfactory
maintenance schedule, it will result in device failure and may endanger human health.
11.1 Checking
Check the following basic items before using the monitor:
Check for any mechanical damage.
Check all exposed wires, insertions and accessories.
Check all instrument functions that may be used for patient monitoring and ensure that the
instrument is in good working condition.
If the instrument function has any sign of damage, do not use this monitor for any patient monitoring.
Please contact the hospital's professional maintenance personnel or our customer service personnel.
Every 6-12 months or after each repair, a comprehensive examination must be performed by trained
and qualified technical service personnel, including functional safety checks; the specific inspection
items are as follows: