84 Lumax Technical Manual
Table 25 Lumax Pacing Modes
Mode
Lumax HF
Lumax DR
Lumax VR
DDDR
X
X
N/A
DDIR
X
X
N/A
VDDR
X
X
N/A
VDIR
X
X
N/A
AAIR
X
X
N/A
VVIR
X
X
X
DDD
X
X
N/A
DDI
X
X
N/A
VDD
X
X
N/A
VDI
X
X
N/A
AAI
X
X
N/A
VVI
X
X
X
OFF
X
X
X
The basic rate timer is started by a sensed or paced event. A
sensed event outside of the refractory period inhibits pacing and
resets the lower rate time; in the absence of a sensed event, a
pacing pulse will be delivered at the end of the lower rate interval.
The pacing modes with an “R” indicate rate adaptive pacing
controlled by a motion based capacitive sensor. These modes
are functionally the same as the corresponding non-rate-adaptive
modes, except that the pacing rate is increased based on
physical activity.
2.9.2 Basic Rate
The basic rate is the pacing rate in the absence of a patient’s
intrinsic rhythm. This rate may be independently programmed for
normal and post-shock bradycardia pacing.