18 Lumax Technical Manual
C
AUTION
Programmed Parameters
– Program the device parameters
to appropriate values based on the patient’s specific
arrhythmias and condition.
Programmers
- Use only BIOTRONIK’s ICS
3000
programmer to communicate with the device.
Defibrillation Threshold
- Be aware that the changes in the
patient’s condition, drug regimen, and other factors may
change the defibrillation threshold (DFT) which may result in
non-conversion of the arrhythmia post-operatively.
Successful conversion of ventricular fibrillation or ventricular
tachycardia during arrhythmia conversion testing is no
assurance that conversion will occur post-operatively.
W
ARNING
Unwanted Shocks
– Always program ICD therapy to
OFF
prior to handling the device to prevent the delivery of shocks
to the patient or the person handling the device during the
implant procedure.
4.8 Implant the ICD/CRT-D
The ICD/CRT-D may be placed in the pocket at this time. Place
the device into the pocket with either side facing up (it can be
interrogated and programmed from either side). Carefully coil any
excess lead length beside or above the ICD/CRT-D.
The pacing and sensing functions of the device should be
evaluated. It is also recommended that at least one induction and
device conversion be done prior to closing the pocket. This will
ensure that the lead system has been securely connected to the
device and has not changed position.
Содержание Lumax DR ICD Series
Страница 8: ...vi Lumax Technical Manual...
Страница 60: ...52 Lumax Technical Manual...
Страница 135: ...6 Lumax Technical Manual...
Страница 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Страница 151: ...22 Lumax Technical Manual...
Страница 168: ...Lumax Technical Manual 39...
Страница 170: ...Lumax Technical Manual 41...
Страница 175: ......