24 Lumax Technical Manual
Table 4: Summary of Observations – Tupos LV/ATx
Category
Number
of
Patients
% of
Patients
Number
per patient-
year
Device Related
Infection
1
0.75%
1
0.01
Inductions and
conversions
6
4.51%
6
0.06
Inappropriate sensing
20
15.04%
20
0.20
Symptomatic with
biventricular pacing
2
1.50%
2
0.02
Total
25
18.80%
29
0.28
Total Procedure,
Lead and Device
Related
61
45.86%
77
0.76
Other Medical Related
Non-CHF Cardiac
Symptoms
21
15.79%
21
0.21
Ventricular
arrhythmias
11
8.27%
11
0.11
Other medical
26
19.55%
32
0.31
Atrial arrhythmia
14
10.53%
14
0.14
Dizziness
4
3.01%
4
0.04
Medication
5
3.76%
5
0.05
Worsening CHF
46
34.59%
46
0.45
Total
82
61.65%
133
1.31
Total – All Patients
and Categories
104
78.20%
210
2.06
Number of Patients = 133 Number of Patient-Years = 101.9