
106 Lumax Technical Manual
W
ARNING
Resuscitation Availability
- Do not perform induction testing
unless an alternate source of patient defibrillation such as an
external defibrillator is readily available. In order to implant
the ICD/CRT-D system, it is necessary to induce and convert
the patient’s ventricular tachyarrhythmias.
C
AUTION
Defibrillation Threshold
- Be aware that the changes in the
patient’s condition, drug regimen, and other factors may
change the defibrillation threshold (DFT) which may result in
non-conversion of the arrhythmia post-operatively.
Successful conversion of ventricular fibrillation or ventricular
tachycardia during arrhythmia conversion testing is no
assurance that conversion will occur post-operatively.
2.10.9 Manual ATP
The ICD/CRT-D can deliver a manual ATP on demand through a
programmer command in the EP test menu. To deliver an ATP
sequence, place the wand over the device and select the
Start
ATP
button. A confirmation menu will appear and the
programmed pacing sequence command will be delivered upon
selecting the
OK
button in this screen. Programming of the
manual ATP is similar to the programming available for automatic
ATP therapy as described in
Section 2.8.2
.
2.10.10 Emergency
Shock
The ICD/CRT-D can deliver an emergency maximum energy
shock on demand through a programmer command in the EP test
menu. For emergency shocks, the energy (maximum 30 or 40 J),
polarity (Standard) and waveform (Biphasic) are pre-defined. To
deliver a shock, place the wand over the device and select the
Emergency Shock
button. A confirmation menu will appear and
the shock will be delivered.
Содержание Lumax DR ICD Series
Страница 8: ...vi Lumax Technical Manual...
Страница 60: ...52 Lumax Technical Manual...
Страница 135: ...6 Lumax Technical Manual...
Страница 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Страница 151: ...22 Lumax Technical Manual...
Страница 168: ...Lumax Technical Manual 39...
Страница 170: ...Lumax Technical Manual 41...
Страница 175: ......