Lumax Technical Manual 105
Burst + PES
Induction delivers a programmed number of pacing
stimuli followed by a programmable number (
Number S1
) of
timed extra stimuli. The burst rate (intervals) is independently
programmable, as is the chamber being stimulated (RV, LV or
BiV). The interval between S1s and the remaining programmed
extra stimuli (PES: S1 through S4 possible) are also
programmable.
Shock on T
induction mode allows tachyarrhythmia induction by
means of a timed T wave shock delivered after a series of paced
stimuli.
Energy
of the T wave shock, number of pulses
(
Number S1
) in the pulse train, synchronization interval (
R-S1
)
and the shock
Coupling
interval are all programmable.
2.10.7 Manual Shock
The ICD/CRT-D can deliver a manual shock on demand through
a programmer command in the EP test menu. For manual
shocks, the
energy
,
polarity
and
waveform
are programmable
by the user. To deliver a shock, place the wand over the device
and select the
Start Shock
button. A confirmation menu will
appear and the shock command will be delivered upon selecting
the
OK
button in this screen. After each manual shock, the EP
test screen will display the shock energy, lead impedance and
charge time.
2.10.8 Test Shock
The ICD/CRT-D can deliver a 1 Joule (R-wave synchronous) test
shock on demand through a programmer command in the EP test
menu. This shock is designed to measure the shock impedance
and test the integrity of the shock electrodes of an implanted ICD
lead.
N
OTE
:
When the test shock is administered, VF detection is
automatically enabled.
Содержание Lumax DR ICD Series
Страница 8: ...vi Lumax Technical Manual...
Страница 60: ...52 Lumax Technical Manual...
Страница 135: ...6 Lumax Technical Manual...
Страница 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Страница 151: ...22 Lumax Technical Manual...
Страница 168: ...Lumax Technical Manual 39...
Страница 170: ...Lumax Technical Manual 41...
Страница 175: ......