Lumax Technical Manual 59
•
The T-wave discrimination period (hold of upper
threshold) is decreased to 100 ms, thus eliminating the
Upper Threshold.
•
The decay rate of the Lower Threshold is increased to
0.125 mV every 250 ms.
These adjustments ensure that the threshold reaches the lower
values more quickly in order to assure that all VF signals are
sensed appropriately.
The Enhanced T-Wave Suppression setting is specifically
designed to avoid double counting of each QRS-T complexes
during normal sinus rhythms. With sensitivity programmed to
Enhanced T-Wave Suppression:
•
High pass filtering is increased to reduce low frequency
signal components such as T-waves and respiratory
artifacts.
•
The Upper Threshold is increased to 75% of the
measured R-wave.
•
The Upper Threshold may not retrigger with each sensed
event, it is only triggered when the new sensed R-wave
crosses the 50% point of the previous measured R-wave.
2.2.2 Minimum Right Ventricular Threshold
This parameter limits the minimum sensitivity of the ICD/CRT-D to
a programmable value. Nominally, the minimum threshold is set
to 0.8 mV, but it can be adjusted from 0.5 to 2.5 mV.
2.2.3 Atrial Sensitivity Settings
DR and HF versions only
The primary option for setting the sensitivity of the atrial input
stage is “Standard”. When atrial sensing is active, the sensitivity
is set to “Standard” for most patients, which is designed to adapt
the parameters of the input stage to various signal conditions.
The available settings are described in
Table 20
.
Содержание Lumax DR ICD Series
Страница 8: ...vi Lumax Technical Manual...
Страница 60: ...52 Lumax Technical Manual...
Страница 135: ...6 Lumax Technical Manual...
Страница 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Страница 151: ...22 Lumax Technical Manual...
Страница 168: ...Lumax Technical Manual 39...
Страница 170: ...Lumax Technical Manual 41...
Страница 175: ......