108 Lumax Technical Manual
W
ARNING
The use of Home Monitoring does not replace regular follow-
up examinations. Therefore, when using Home Monitoring,
the time period between follow-up visits may not be extended.
The implant’s Home Monitoring function can be used for the
entire operational life of the implant (prior to EOS) or for shorter
periods, such as several weeks or months. Home Monitoring is
programmable, ON or OFF.
N
OTE
:
When ERI mode is reached, this status is transmitted. All
measurements and transmissions are unaffected. When End
of Service (EOS) is reached and transmitted, further
measurements and transmissions of Home Monitoring data
are no longer possible.
2.11.2.1 Transmission of Information
The implant transmits information with a small transmitter, which
has a range of about 2 meters. The patient’s implant data are
sent to the corresponding patient device in configurable periodic
intervals. The transmissions may also be activated by the
detection of a cardiac event, as programmed. The types of
transmissions are discussed in
Section 2.11.2.4
.
The minimal distance between the implant and the patient device
must be maintained at 15 cm.
2.11.2.2 CardioMessenger
The CardioMessenger II patient device (
Figure 10
) is designed
for use in the home and is comprised of the mobile device and the
associated charging station. The patient can carry the mobile
device with them during his or her occupational and leisure
activities. The patient device is rechargeable, allowing for an
approximate operational time of approximately 24 hours. It
receives information from the implant and forwards it via
telephone networks (cellular or traditional) to the BIOTRONIK
Service Center.
Содержание Lumax DR ICD Series
Страница 8: ...vi Lumax Technical Manual...
Страница 60: ...52 Lumax Technical Manual...
Страница 135: ...6 Lumax Technical Manual...
Страница 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Страница 151: ...22 Lumax Technical Manual...
Страница 168: ...Lumax Technical Manual 39...
Страница 170: ...Lumax Technical Manual 41...
Страница 175: ......