Lumax Technical Manual 9
W
ARNING
ICD Lead Systems
- BIOTRONIK ICDs/CRT-Ds maybe
implanted with any legally marketed, compatible ICD lead.
Compatibility is defined as:
•
IS-1 pacing and sensing connector(s)
•
DF-1 shock coil connector(s)
•
Integrated or dedicated bipolar pacing and sensing
configuration
•
Active or passive fixation technology
•
Single or dual defibrillation shock coil (s)
•
High energy shock accommodation of at least 30
joules
•
Insertion and withdrawal forces as specified by
ISO 5841-3 (IS-1) and ISO 11318:1993 (E) DF-1
The following leads were evaluated in a retrospective study
with BIOTRONIK’s ICDs/CRT-Ds:
•
Medtronic Sprint 6932
•
Medtronic Sprint 6943
•
Medtronic Sprint Quattro 6944
•
Medtronic Transvene RV 6936
•
St. Jude (Ventritex) TVL- ADX 1559
•
St. Jude SPL SP02
•
Guidant Endotak DSP
•
Guidant Endotak Endurance EZ, Endotak Reliance
•
Guidant (Intermedics) 497-24.
The following leads were bench tested for compatibility with
BIOTRONIK’s ICDs/CRT-Ds:
•
Guidant Endotak Endurance “CPI 0125”
•
Guidant Endotak Reliance 0148
•
Medtronic Sprint 6932
•
Medtronic Sprint 6942
•
Medtronic Sprint 6943
•
Medtronic Sprint 6945
•
Medtronic Sprint Quattro 6944
•
St. Jude Riata 1571/65
•
St. Jude SPL SPO1.”