![BIOTRONIK ICD Series Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/biotronik/icd-series/icd-series_technical-manual_2756135042.webp)
40
Technical Data
Electrical Characteristics
Standards
The specifications are made according to EN 45502-2-2:2008.
Measuring conditions
If not indicated otherwise, all specifications refer to the following conditions:
•
Ambient temperature: 37 ºC ± 2 °C
•
Pacing/sensing: 500 Ω ± 1%
•
Shock: 50 Ω ±1%
Factory settings
•
Arrhythmia zones VT1, VT2, VF: OFF
•
Antibradycardia pacing: OFF
•
Home Monitoring: OFF
Telemetry data
•
Nominal carrier frequency: 403.6 MHz
•
Maximum power of transmission: < 25 µW (-16 dBm)
International radio certifi-
cation
Devices with BIOTRONIK Home Monitoring
®
are equipped with an antenna for
wireless communication.
Telemetry data for Canada and the USA:
This device must neither interfere with meteorological and earth resources tech-
nology satellites nor with meteorological stations working in the 400,150 to 406,000
MHZ band, and it must accept any interference received, including interference
that may cause undesired operation.
•
This device will be registered with Industry Canada under the following
number:
IC: 4708A-TACHNXT
The code IC in front of the certification/ registration number only indicates that
the technical requirements for Industry Canada are met.
•
This device will be registered with Federal Communications Commission under
the following number:
FCC ID: QRITACHNXT
Telemetry data for Japan:
In accordance with Japanese law, this device has been assigned an identification
number under the "Ordinance concerning certification of conformity with technical
regulations etc. of specified radio equipment", Article 2-1-8.
•
R: 202-SMA026
Pulse form
The pacing pulse has the following form:
The pulse amplitude reaches its maximum value at
the beginning of the pulse (Ua). With increasing
pacing duration (tb), the pulse amplitude is
reduced dependent on the pacing impedance.