32
Parameters
Timing: Rate adaptation via
accelerometer
Pacing: Pulse amplitude
and pulse width
Pacing: Ventricular capture
control
Pacing: atrial capture
control
LV lead configuration
Parameter
Range of values
Standard
VR
DX
DR
HF
Maximum sensor rate
80 ... (10) ... 160 bpm
160 bpm
x
x
x
x
Sensor gain
AUTO
Very low; Low; Medium; High;
Very high
Medium
x
x
x
x
Sensor threshold
Very low; Low; Medium; High;
Very high
Medium
x
x
x
x
Rate increase
1; 2; 4; 8 bpm/cycle
2 bpm
x
x
x
x
Rate decrease
0.1; 0.2; 0.5; 1.0 bpm/cycle
0.5 bpm
x
x
x
x
Parameter
Range of values
Standard
VR
DX
DR
HF
Pulse amplitude A
0.5 ... (0.25) ... 4.0 ... (0.5) ... 6.0; 7.5 V
2.5 V
x
x
Pulse amplitude V/RV
x
x
x
x
Pulse amplitude LV
x
Pulse width A
0.4; 0.5 ... (0.25) ... 1.5 ms
0.4 ms
x
x
Pulse width V/RV
x
x
x
x
Pulse width LV
0.5 ms
x
Parameter
Range of values
Standard
VR
DX
DR
HF
Capture control
OFF; ATM; ON
ATM
x
x
x
x
Threshold test start
2.5 ... (0.5) ... 5.0 V
ATM: 2.5 V
ON: 3.5 V
x
x
x
x
Minimum amplitude
1.0 ... (0.25) ... 4.0 V
1.0 V
x
x
x
x
Safety margin
1.0; 1.2 V
1.0 V
x
x
x
x
Parameter
Range of values
Standard
VR
DX
DR
HF
Capture control
OFF; ATM
ATM
x
x
Parameter
Range of values
Standard
VR
DX
DR
HF
LV pacing polarity
LV tip -> LV ring;
LV tip -> RV ring;
LV ring -> LV tip;
LV ring -> RV ring;
UNIP
LV tip -> RV
ring
x
LV sensing polarity
UNIP; BIPL
UNIP
x