20
Implantation
Implanting
Applying the programming
head
The programming head (PGH) features a diagram of the device. This is used to
assist in positioning the head to ensure proper telemetry.
•
Make sure the PGH is positioned correctly.
Establishing telemetry
contact
The programmer (or the SafeSync Module) can be no more than 3 m from the
device; ideally there should be no hindrances between the patient and the pro-
grammer.
•
Switch on RF telemetry on the programmer.
•
Apply the programming head for about 2 s until successful initialization is dis-
played on the programmer:
The SafeSync symbol is displayed in the navigator and the signal
strength is displayed in the status line.
•Remove the programming head.
Activating ICD therapy
•
Load the device program that is suitable for the device type in the programmer.
•
Activate ICD therapy.
•
Shipment mode is permanently deactivated once the leads have been con-
nected and initial measurement of the pacing impedance has been performed.
The device data are saved.
•
Take precautionary measures while programming.
•
If the device induces tachycardia while programming ATPs or does not deliver
adequate therapy in the DFT test: use emergency shock or an external defibril-
lator.
1
Prepare the vein.
2
Implant the leads, perform the measurements, and fixate the leads.
3
Form the device pocket.
4
Connect the lead connector to the device.
5
Insert the device.
6
Guide the fixation suture through the opening in the header and fixate the
device in the prepared device pocket.
7
Close the device pocket.
8
Check the device with standard tests.