23
Implantation
Considering power
consumption and service
time
RF telemetry requires somewhat more power: Consumption during implantation
corresponds to approximately 10 days of service time and consumption during a
20-minute follow-up corresponds to approximately 3 days.
•
Do not establish unnecessary RF telemetry.
•
After 5 minutes without input, SafeSync switches to the economy mode.
•
Check the battery capacity of the device at regular intervals.