Parameters
Sensing
Sensitivity and thresholds
Parameter
Range of values
Standard
VR
DX
DR
HF
Sensing A
STD; OFF; IND
STD
x
x
x
Sensing RV
STD; TWS; VFS; IND
STD
x
x
x
x
Sensing LV
STD; OFF; IND
STD
x
Upper threshold RV
50; 75 %
50%
x
x
x
x
Upper threshold LV
50; 75 %
50%
x
Upper threshold duration
after detection
110; 150 ... (50) ... 500 ms
VFS: 110 ms
350 ms
x
x
x
x
Upper threshold duration after pacing
400 ms
Lower threshold RV
25; 50%
25%
x
x
x
x
T-wave suppression after pacing
OFF; ON
OFF
x
x
x
x
Minimum threshold A
0.2 ... (0.1) ... 2.0 mv
0.4 mv
x
x
x
Minimum threshold RV
0.5 ... (0.1) ... 2.5 mv
0.8 mv
x
x
x
x
Minimum threshold LV
0.5 ... (0.1) ... 2.5 ... (0.5) ... 5.0 mv
1.6 mv
x