19
Implantation
Connecting the lead
connector to the device
Keeping distance between
leads
1
Disconnect stylets and stylet guides.
2
DF-1/IS-1 connection:
•
Connect the DF-1 connector for the right-ventricular shock coil to RV.
•
Connect the DF-1 connector for the supraventricular shock coil to SVC.
Or connect a subcutaneous array to SVC.
DF4/IS-1 connection:
•
Connect the DF4 connector to RV
3
DF-1/IS-1 connection:
•
Connect the bipolar IS-1 lead connector for the atrium to RA.
•
Connect the IS-1 lead connector for the right ventricle to RV.
•
Connect the unipolar or the bipolar IS-1 lead connector for the left ven-
tricle to LV.
DF4/IS-1 connection:
•
Connect the bipolar IS-1 lead connector for the atrium to RA.
•
Connect the unipolar or the bipolar IS-1 lead connector for the left ven-
tricle to LV.
4
Push the lead connector into the header without twisting or bending the
connector or conductor until the connector tip (on the DF-1 connector) or
the insertion indicator (on the DF4 connector) becomes visible behind the
set screw block. This indicator can vary depending on the manufacturer of
the lead used.
5
If you cannot easily plug the lead connector into the connection:
•
Use only sterile water as lubricant.
6
If the lead connector cannot be inserted completely, the set screw may be
protruding into the drill hole of the set screw block.
•
Use the screwdriver to perpendicularly pierce through the slitted point
in the center of the silicone plug until it reaches the set screw.
•
Carefully loosen the set screw without completely unscrewing it, so that
it does not become tilted upon retightening.
7
Turn the set screw clockwise until torque control starts (you will hear a
clicking sound).
8
Carefully withdraw the screwdriver without retracting the set screw.
•
In case of IS-1 connections with two set screws, tighten both screws!
•
When the screwdriver is withdrawn, the silicone plug automatically
safely seals the lead connector port.
!
!
WARNING
Inadequate therapy
When leads are not spaced sufficiently apart or are positioned inappropriately,
this can lead to far-field sensing or insufficient defibrillation.
•
The distance between 2 shock coils must be greater than 6 cm.
•
Tip and ring electrodes must not have contact with each other.