18
Implantation
Checking parts
Damage to any of the parts can result in complications or technical failures.
•
Check for damage before and after unpacking all parts.
•
Replace damaged parts.
•
The ICD is shipped with tachyarrhythmia therapy deactivated and is only to be
connected and implanted in this state.
•
Leads may not be shortened.
Implantation site
•
Depending on lead configuration and the patient's anatomy, the ICD is generally
implanted subpectorally on the left side.
Preventing leakage
currents
Leakage currents between the tools and the device must be prevented during
implantation.
•
Electrically insulate the patient.
Preventing unintentional
shock delivery
Avoiding damage to the
header
There is a blind plug for DF-1 and IS-1 connections in the header. The provided set
screws must be carefully loosened or tightened.
•
Loosen set screws with the supplied screwdriver. Use only BIOTRONIK screw-
drivers with torque control!
•
Do not forcibly pull out the blind plug!
•
If lead repositioning is necessary, re-order sterile screwdrivers from BIO-
TRONIK.
Preventing short circuits in
the header
Ensure that connections are
clean
In case of contamination during implantation:
•
Clean lead connectors with a sterile cloth.
•
Rinse connection only with sterile water.
!
!
WARNING
Shock delivery with activated ICD
There is a risk of unintended shock delivery when handling an activated ICD.
•
Deactivate ICD therapy before touching the device during implantation,
device replacement and explantation.
!
!
WARNING
Short circuit due to open lead connector ports
Connector ports in the header which are open and thus not electrolyte-proof may
cause undesired current flows to the body and penetration of body fluid into the
device.
•
Either leave unused ports closed with the premounted blind plugs, or close
them using the supplied blind plugs.