![BIOTRONIK ICD Series Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/biotronik/icd-series/icd-series_technical-manual_2756135027.webp)
25
Implantation
Follow-up
Follow-up intervals
Follow-ups must be performed at regular, agreed intervals.
•
The first follow-up should be carried out by the physician using the pro-
grammer (in-office follow-up) approximately 3 months after implantation fol-
lowing the lead ingrowth phase.
•
The next in-office follow-up should be carried out once a year and no later than
12 months after the last in-office follow-up.
Follow-up with BIOTRONIK
Home Monitoring
®
Monitoring using the Home Monitoring function does not serve to replace regular
in-office appointments with the physician required for other medical reasons.
Follow-up supported by Home Monitoring can be used to functionally replace in-
office follow-up under the following conditions:
•
The patient was informed that the physician must be contacted despite use of
the Home Monitoring function if symptoms worsen or if new symptoms arise.
•
Device messages are transmitted regularly.
•
The physician decides whether the data transmitted via Home Monitoring with
regard to the patient's clinical condition as well as the technical state of the
device system are sufficient. If not, an in-office follow-up has to be carried out.
Possible early detection due to information gained via Home Monitoring may
necessitate an additional in-office follow-up. For example, the data may indicate at
an early stage lead problems or a foreseeable end of service time (ERI). Further-
more, the data could provide indications of previously unrecognized arrhythmias or
modification of the therapy by reprogramming the device.
Follow-up with the
programmer
Use the following procedure for in-office follow-up:
1
Record and evaluate the ECG.
2
Interrogate the device.
3
Evaluate the status and automatically measured follow-up data.
4
Check the sensing and pacing functions.
5
Possibly evaluate statistics and IEGM recording.
6
Manually perform standard tests if necessary.
7
Possibly customize program functions and parameters.
8
Transmit the program permanently to the device.
9
Print and document follow-up data (print report).
10
Finish the follow-up for this patient.
Содержание ICD Series
Страница 4: ...2 ...
Страница 6: ...4 Table of Contents ...
Страница 40: ...38 Parameters ...
Страница 48: ...46 Technical Data Bipolar IS 1 connector Unipolar IS 1 connector Unipolar DF 1 connector DF4 connector ...