Chapter 6
Statistics
Evia HF / HF-T Technical Manual
PAGE 81
Figure 51: Far-Field Histogram
The Far-field histogram in
provides information related to sensed events in the atrium
occurring after ventricular paced or sensed events. Data is measured in 10 ms intervals. The maximum
value measured is 225 ms after a ventricular paced or sensed event.
The area in gray represents the programmed interval range of the far-field protection parameter. Events
outside the programmed far-field protection interval will be shown in the white portion of the graph.
6.6.3 Ap-Vs interval distribution curve
This graph provides information related to the amount of Vs response to atrial paced events. The
information is divided into 5 rate bins and provides the minimum, mean and maximum Ap-Vs intervals
for each rate bin. The programmed AV Delay is also shown.
The data is also displayed on a graph with the Y axis showing the range of programmable AV Delay
options and the X axis graph showing heart rate.
The number of successful AV hysteresis scans is provided on this graph.
Figure 52: Ap-Vs Interval Distribution
Содержание Evia HF
Страница 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Страница 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Страница 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Страница 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Страница 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Страница 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Страница 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Страница 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Страница 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Страница 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...