Chapter 5
Programmable Parameters
Evia HF / HF-T Technical Manual
PAGE 64
• Mode Switching
• Number of Mode Switches
• Duration of Mode Switches
Ventricular Arrhythmia
• High Ventricular Rate Counters
Transmitted Device Settings
The primary programmed parameters for the following are sent in the data package:
• Leads—(e.g., Pacing Output, Configuration)
• Brady—(e.g., Basic Rate, UTR, AV-Delays, RV Sensitivity)
• I-OPT—(ON/OFF)
• HM Settings—(e.g., ON/OFF, transmission time (daily), IEGM transmissions ON/OFF, periodic
IEGM, ongoing atrial episode, statistics, holter)
• Miscellaneous other information
5.18.5.1 IEGM Online HDs
The Evia HF-T provides the ability to transmit periodic IEGM Online HD (IEGM and marker data) from
the periodic follow-ups as an addition to the current messages.
An IEGM with up to 3 channels (LV, RV and/or RA) are sent in one message.
The following markers are also transmitted: A
S
(including A
rs
), A
S
(PMT), A
P
, RV
S
/LV
S
(including RV
rs
)
and RV
P
/LV
P
.
The Evia HF-T includes a programmable parameter to disable or enable the IEGM transmission.
Содержание Evia HF
Страница 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Страница 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Страница 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Страница 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Страница 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Страница 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Страница 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Страница 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Страница 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Страница 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...