5.5.4 AV Delay
Maximum AV Delay Programmable Value
350 ms
Maximum AV Delay Programmable Value
with AV Hysteresis
450 ms
I-Opt AV Delay maximum
3
400 ms
Lower Rate AV Delay Upper Rate AV Delay
Low
150 ms
120 ms
Medium
150 ms
75 ms
High
150 ms
50 ms
4
Fixed
150 ms
150 ms
Individual
User Defined
User Defined
AV Delay Settings (Default for BiV programming)
Evia CRT-P provides three preset AV Delay settings: Low, Medium and High. In addition, the user
can program a fixed AV Delay, as well as a user-defined program (Individual). Similar to previous
generations of devices, Evia CRT-P allows the user to program individual AV Delay programs. Evia
CRT-P adds the option of changing the values on the AV Delay visual display in addition to changing
the numeric parameter buttons. By placing the pen over the “o” on the display, the user can slide the
value up or down to increase or decrease the AV Delay setting for a particular rate bin.
5.5.4.1 Dynamic AV Delay
The AV delay defines the interval between an atrial paced or sensed event and the ventricular pacing
pulse. If the pulse generator is programmed to a dual chamber sensing mode, an intrinsic ventricular
event falling within the AV delay will inhibit the ventricular pacing pulse. If not contraindicated, a longer
AV delay can be selected to increase the probability of ventricular output pulse inhibition. Short AV
delays are available for testing purposes or if ventricular pre-excitation is desired (i.e., hemodynamic
considerations).
Dynamic AV Delay provides independent selection of AV Delays from five rate ranges at pre-set AV
Delay values. In addition, the AV Delay after atrial pace events can be differentiated from the AV interval
after atrial sense events for dual chamber pacing modes. Dynamic AV Delay is programmable within
the following atrial rate ranges at the values specified in
3 I-Opt available in the Evia HF-T if RV only pacing if programmed
4 Conflict with Sense compensation will arise as a minimum of 15 ms AV Delay is required
Содержание Evia HF
Страница 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Страница 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Страница 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Страница 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Страница 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Страница 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Страница 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Страница 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Страница 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Страница 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...